FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 2011943 · Received February 14, 2011

Report

Report Number
3002158293-2011-00138
Event Type
Malfunction
Date Received
February 14, 2011
Date of Event
December 16, 2010
Report Date
February 5, 2011
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF CHARGER/MODEM (B)(4) HAS BEEN COMPLETED. AS RECEIVED, THE CHARGER/MODEM WAS FOUND TO HAVE DEFECTIVE COMPONENTS. THE FLASH MEMORY CHIPS (COMPONENTS U102 AND U105) WERE CORRUPT AND NEEDED TO BE REPLACED. THE ROOT CAUSE OF THE DEFECTIVE FLASH MEMORY CANNOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENTS RESULTED FROM THE DEFECTIVE CHARGER/MODEM.

Description of Event or Problem · 1

A REVIEW OF SERVICE DATA DETECTED A REPORTABLE EVENT. DURING SERVICING, CHARGER/MODEM (B)(4) WAS UNABLE TO POWER ON. THE LAST PATIENT TO USE THIS CHARGER/MODEM DID NOT REPORT ANY DEFICIENCIES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 NA