LIFEVEST WCD 4000 SYSTEM
MDR report key: 2011943
·
Received February 14, 2011
Report
- Report Number
- 3002158293-2011-00138
- Event Type
- Malfunction
- Date Received
- February 14, 2011
- Date of Event
- December 16, 2010
- Report Date
- February 5, 2011
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF CHARGER/MODEM (B)(4) HAS BEEN COMPLETED. AS RECEIVED, THE CHARGER/MODEM WAS FOUND TO HAVE DEFECTIVE COMPONENTS. THE FLASH MEMORY CHIPS (COMPONENTS U102 AND U105) WERE CORRUPT AND NEEDED TO BE REPLACED. THE ROOT CAUSE OF THE DEFECTIVE FLASH MEMORY CANNOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENTS RESULTED FROM THE DEFECTIVE CHARGER/MODEM.
Description of Event or Problem · 1
A REVIEW OF SERVICE DATA DETECTED A REPORTABLE EVENT. DURING SERVICING, CHARGER/MODEM (B)(4) WAS UNABLE TO POWER ON. THE LAST PATIENT TO USE THIS CHARGER/MODEM DID NOT REPORT ANY DEFICIENCIES.