FDA Adverse Event Death Summary report: N

VERSACARE BED

MDR report key: 2011864 · Received March 4, 2011

Report

Report Number
1824206-2011-01252
Event Type
Death
Date Received
March 4, 2011
Date of Event
January 28, 2011
Report Date
February 8, 2011
Manufacturer
HILL-ROM INC.
Product Code
FNL
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN STATED THAT HE TESTED THE PT POSITIONING MONITOR (PPM) SYSTEM FOR THE BED AND FOUND IT TO BE WORKING PROPERLY. THE TECHNICIAN FOUND THE BED SCALE WAS ZEROED WHILE THE PT WAS IN THE BED AND THE BED EXIT WAS NOT SET. THE TECHNICIAN ALSO DISCOVERED THAT A SIDECOMM COMMUNICATION CABLE WAS NOT CONNECTED FROM THE BED TO THE ACCOUNTS NURSE CALL SYSTEM. ADDITIONALLY THE TECHNICIAN INSPECTED ALL OF THE SIDERAILS AND FOUND THEM TO BE OPERATING PROPERLY. THE ACCOUNT WOULD NOT RELEASE INFO REGARDING THE PT OR IF THE PT DIED AS A RESULT OF THE FALL. VERSACARE BED USER MANUAL STATES: THE BED EXIT ALARM SYSTEM SHOULD BE ZEROED BEFORE PUTTING THE PT ON THE BED. VERSACARE BED USER MANUAL STATES: THE BED EXIT DOES NOT ALARM AND ALL THREE MODE INDICATORS ARE FLASHING, REMOVE THE PT AND ZERO THE BED EXIT SYSTEM. VERSACARE BED USER MANUAL STATES: WARNING: A COMMUNICATION CABLE MUST BE USED FOR BEDS THAT HAVE NURSE CALL. FAILURE TO DO COULD CAUSE PT INJURY. FOR BEDS THAT HAVE NURSE CALL SYSTEMS, A COMMUNICATION CABLE MUST BE CONNECTED BETWEEN THE BED AND FACILITY COMMUNICATION SYSTEM.

Description of Event or Problem · 1

THE DIRECTOR OF PT SERVICES ALLEGED THAT ON (B)(6), 2011, A PT FELL OUT OF BED AND WAS FOUND ON THE FLOOR AND THE PT EXPIRED. THERE WERE NO WITNESSES. THE ATTENDING NURSE TOLD THE DIRECTOR THAT THE BED EXIT ON THE BED DID NOT ALARM AND THREE BED EXIT LIGHTS WERE FLASHING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VERSACARE BED AC POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC. 3200

Patients

Seq Age Sex Outcome Treatment
1 UNK Death