VERSACARE BED
Report
- Report Number
- 1824206-2011-01252
- Event Type
- Death
- Date Received
- March 4, 2011
- Date of Event
- January 28, 2011
- Report Date
- February 8, 2011
- Manufacturer
- HILL-ROM INC.
- Product Code
- FNL
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER
Narratives
THE TECHNICIAN STATED THAT HE TESTED THE PT POSITIONING MONITOR (PPM) SYSTEM FOR THE BED AND FOUND IT TO BE WORKING PROPERLY. THE TECHNICIAN FOUND THE BED SCALE WAS ZEROED WHILE THE PT WAS IN THE BED AND THE BED EXIT WAS NOT SET. THE TECHNICIAN ALSO DISCOVERED THAT A SIDECOMM COMMUNICATION CABLE WAS NOT CONNECTED FROM THE BED TO THE ACCOUNTS NURSE CALL SYSTEM. ADDITIONALLY THE TECHNICIAN INSPECTED ALL OF THE SIDERAILS AND FOUND THEM TO BE OPERATING PROPERLY. THE ACCOUNT WOULD NOT RELEASE INFO REGARDING THE PT OR IF THE PT DIED AS A RESULT OF THE FALL. VERSACARE BED USER MANUAL STATES: THE BED EXIT ALARM SYSTEM SHOULD BE ZEROED BEFORE PUTTING THE PT ON THE BED. VERSACARE BED USER MANUAL STATES: THE BED EXIT DOES NOT ALARM AND ALL THREE MODE INDICATORS ARE FLASHING, REMOVE THE PT AND ZERO THE BED EXIT SYSTEM. VERSACARE BED USER MANUAL STATES: WARNING: A COMMUNICATION CABLE MUST BE USED FOR BEDS THAT HAVE NURSE CALL. FAILURE TO DO COULD CAUSE PT INJURY. FOR BEDS THAT HAVE NURSE CALL SYSTEMS, A COMMUNICATION CABLE MUST BE CONNECTED BETWEEN THE BED AND FACILITY COMMUNICATION SYSTEM.
THE DIRECTOR OF PT SERVICES ALLEGED THAT ON (B)(6), 2011, A PT FELL OUT OF BED AND WAS FOUND ON THE FLOOR AND THE PT EXPIRED. THERE WERE NO WITNESSES. THE ATTENDING NURSE TOLD THE DIRECTOR THAT THE BED EXIT ON THE BED DID NOT ALARM AND THREE BED EXIT LIGHTS WERE FLASHING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VERSACARE BED | AC POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM INC. | 3200 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Death |