FDA Adverse Event Malfunction Summary report: N

6461500 INFUSOR PUMP (1 LABEL)

MDR report key: 2011693 · Received March 9, 2011

Report

Report Number
6000001-2011-01727
Event Type
Malfunction
Date Received
March 9, 2011
Date of Event
February 16, 2011
Report Date
February 16, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K883577
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). A SERVICE HISTORY REVIEW WAS PERFORMED REVEALING THIS PUMP HAS NEVER PREVIOUSLY BEEN SENT TO BAXTER FOR SERVICE AND, THEREFORE, THE REPORTED CONDITION IS NOT RELATED TO ANY PREVIOUS REPORTED PROBLEM FOR THIS PUMP. A DEVICE HISTORY REVIEW WAS ALSO PERFORMED, FINDING THAT NO EXCEPTIONS WERE NOTED DURING THE MANUFACTURING OF THIS DEVICE. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THIS DEVICE WAS RETURNED TO BAXTER FOR EVALUATION. A VISUAL INSPECTION AND FUNCTIONAL TESTS WERE PERFORMED. DEVICE EVALUATION CONFIRMED THE REPORTED CONDITION OF A CONSTANT ALARM. THE ROOT CAUSE WAS DETERMINED TO BE A MALFUNCTIONING MICRO PROCESSING UNIT PRINTED CIRCUIT BOARD. THE MICRO PROCESSING UNIT PRINTED CIRCUIT BOARD WAS REPLACED TO REPAIR THE REPORTED CONDITION. A FOLLOW UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED AN INFUSO.R. DEVICE WHICH WAS ALARMING AND BEEPING IN THE ANESTHESIA DEPARTMENT. THIS CONDITION OCCURRED DURING DELIVERY AND MAY HAVE INTERRUPTED DEVICE DELIVERY. DEVICE EVALUATION CONFIRMED THIS CONDITION AS A CONSTANT ALARM WHICH INTERRUPTED DELIVERY. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 6461500 INFUSOR PUMP (1 LABEL) PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1