COLLEAGUE COLOR PUMP CE FRENCHVERSION SINGLE
Report
- Report Number
- 6000001-2011-01719
- Event Type
- Malfunction
- Date Received
- March 9, 2011
- Date of Event
- February 13, 2011
- Report Date
- February 15, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- Removal / Correction Number
- 6000001-3/15/05-007-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH FAILURE CODE 810:11 WAS CONFIRMED BY SERVICE BUT NOT DUPLICATED. THE REPORTED PROBLEM WAS ATTRIBUTED TO MOISTURE IN THE TUBING CHANNEL; NO REPAIR WAS REQUIRED BUT THE TUBING CHANNEL WAS CLEANED AND DRIED AS A PRECAUTION. ADDITIONAL INFORMATION: A DEVICE HISTORY REVIEW WAS PERFORMED FINDING NO EXCEPTION, NONCONFORMANCE, OR REWORK THAT OCCURRED DURING THE MANUFACTURING OF THE DEVICE WITH THIS SERIAL NUMBER. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH CAPA MDQ-CAPA (B)(4).
(B)(4). THE DEVICE IS CURRENTLY IN THE PROCESS OF BEING EVALUATED ON SITE BY A BAXTER FIELD SERVICE TECHNICIAN. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE. THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT CONTAIN A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US.
IT WAS REPORTED TO BAXTER (B)(4) THAT A COLLEAGUE INFUSION PUMP EXPERIENCED FAILURE CODE 810:11. IT IS UNKNOWN WHEN OR IN WHICH CARE AREA THIS EVENT OCCURRED. UPON REVIEWING THE PUMP'S EVENT HISTORY, BAXTER QUALITY ENGINEERING DISCOVERED THAT THIS EVENT INTERRUPTED DELIVERY. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 6.63.92.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE COLOR PUMP CE FRENCHVERSION SINGLE | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |