PTERYFIT IMPLANT
Report
- Report Number
- 3011390931-2024-00642
- Event Type
- Injury
- Date Received
- September 1, 2024
- Date of Event
- March 11, 2024
- Report Date
- September 1, 2024
- Manufacturer
- NORIS MEDICAL LTD
- Product Code
- DZE
- UDI-DI
- 07290108817133
- PMA / PMN Number
- K140440
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
UDI RELATED DATA QUALITY UPDATES ONLY.
THE DHR WAS REVIEWED FOR LOT# 9000649 AND THERE WAS NO EVIDENCE DISCOVERED TO INDICATE THAT A PRODUCT DEFECT OR NON-CONFORMITY CONTRIBUTED TO THE ISSUE. THE INVESTIGATION FOR THIS EVENT HAS NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. OUR MANUFACTURING SYSTEM ASSURES, THAT PRODUCTION AND PROCESS CONTROLS ARE IN PLACE TO ENSURE THAT BATCHES CONFORM TO THE APPLICABLE SPECIFICATIONS BEFORE THEY ARE DISTRIBUTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.
A HEALTHCARE PROFESSIONAL REPORTED THAT NMAF4225 LOT# 9000649 IMPLANT LACKED PRIMARY STABILITY ON TOOTH #2. THE IMPLANT WAS REMOVED DURING IMPLANT PLACEMENT WITH NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS. A SECONDARY IMPLANT WAS PLACED AFTER REMOVAL. ACCORDING TO THE INFORMATION, THE PATIENT IS HEALTHY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO OTHER MEDICAL ISSUES WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 804414 | PTERYFIT IMPLANT | PTERYFIT DENTAL IMPLANT | DZE | NORIS MEDICAL LTD | NMAF4225 | 9000649 | 07290108817133 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Female |