PUMP,COLLEAGUE CXE SINGLECHANNEL COLOUR FRENCH LOANER
Report
- Report Number
- 6000001-2011-01709
- Event Type
- Malfunction
- Date Received
- March 9, 2011
- Date of Event
- February 1, 2011
- Report Date
- February 14, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- Removal / Correction Number
- 6000001-12/1/08-002-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). A 510(K) NUMBER WILL NOT BE PROVIDED IN THE EMDR AS THIS PRODUCT IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE U.S. AND DOES NOT HAVE A 510(K) NUMBER. HOWEVER, IT IS BEING REPORTED BECAUSE IT IS SAME AS OR SIMILAR TO PRODUCT DISTRIBUTED WITHIN THE U.S. REVIEW OF THE DEVICE EVENT HISTORY DETERMINED THAT THE REPORTED CONDITION OCCURRED ON (B)(4), 2011 AND NOT THE ORIGINALLY REPORTED OCCURRENCE DATE OF (B)(4), 2011. DEVICE EVALUATION: THIS DEVICE WAS NOT RETURNED TO BAXTER FOR EVALUATION, HOWEVER, A BAXTER FIELD SERVICE TECHNICIAN REPAIRED THIS DEVICE AT THE CUSTOMER SITE. A VISUAL INSPECTION AND FUNCTIONAL TESTS WERE PERFORMED. DEVICE EVALUATION CONFIRMED THE REPORTED CONDITION OF A BATTERY ISSUE. THE ROOT CAUSE WAS DETERMINED TO BE DEPLETED MAIN BATTERIES. THE MAIN BATTERIES AND BATTERY HARNESS WERE REPLACED TO REPAIR THE REPORTED CONDITION. A FOLLOW UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
(B)(4). ADDITIONAL INFORMATION: BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4). THIS INVOLVED A COLLEAGUE P1.5 INFUSION PUMP WITH A USER INTERFACE MODULE MASTER SOFTWARE VERSION 6.13.92. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.
ON (B)(6) 2011, THE BAXTER FIELD SERVICE TECHNICIAN SERVICED A COLLEAGUE DEVICE, PRODUCT CODE DNM9161L, SN (B)(4), FOR BATTERY ISSUE. PATIENT INJURY WAS NOT REPORTED. THE BAXTER FIELD SERVICE TECHNICIAN REPAIRED THE DEVICE ON SITE. AS SUCH, THE DEVICE WILL NOT BE RETURNED TO CANADIAN TECHNICAL SERVICES. INCIDENT DATE: UNKNOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PUMP,COLLEAGUE CXE SINGLECHANNEL COLOUR FRENCH LOANER | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |