FDA Adverse Event Injury Summary report: N

MINIMED QUICK-SET

MDR report key: 20114021 · Received August 31, 2024

Report

Report Number
3003442380-2024-23464
Event Type
Injury
Date Received
August 31, 2024
Date of Event
July 31, 2024
Report Date
July 29, 2025
Manufacturer
UNOMEDICAL A/S
Product Code
FPA
UDI-DI
05705244017573
PMA / PMN Number
K160648
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

E1: (B)(6).

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS - H11: INVESTIGATION SUMMARY: THE INFORMATION IN THIS COMPLAINT (B)(4) HAS BEEN EVALUATED FOR THE MALFUNCTION CODE NO MALFUNCTION BASED ON COMPLAINT INFORMATION. THE BATCH 6004005 IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. TEST RESULTS: COMPLAINT INVESTIGATION: THE REFERENCE SAMPLES CANNOT BE TESTED BECAUSE THERE WAS NO SPECIFIC MALFUNCTION TO INVESTIGATE. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6004005 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 77 ON 05-NOV-2023, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED. TRENDING: A QUERY WAS RUN IN DATABASE ON 28-JUL-2025 AGAINST HARM CODE SIGNS OF UNTREATED HYPOGLYCAEMIA THAT PATIENT IS UNABLE TO SELF-MANAGE REQUIRING INTERVENTION BY AN HEALTH CARE PROFESSIONAL (HCP) OR REQUIRES EMERGENCY ADVANCED LIFE SUPPORT TO PREVENT PERMANENT ORGAN DAMAGE, MALFUNCTION CODE NO MALFUNCTION BASED ON COMPLAINT INFORMATION DESCRIBES AND LOT 6004005 AND NO MORE COMPLAINT HAS BEEN REGISTERED IN DATABASE FOR THE SAME LOT, HARM CODE AND MALFUNCTION CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION, ONLY ONE COMPLAINT RECEIVED ON THE LOT IN QUESTION AND HARM CODE, THE TEST ON REFERENCE SAMPLES WERE NOT TESTED SINCE A NO MALFUNCTION RELATED TO THE INFUSION SET WAS REPORTED, THEREFORE NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CAPA PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. ON 31-JULY-2024, IT WAS REPORTED BY THE PATIENT THAT HE WAS HOSPITALIZED FOR 2 HOURS DUE TO HYPOGLYCEMIA. PATIENT WAS SLEEPING WHEN HIS LEGS STARTED SHAKING. LOW BLOOD GLUCOSE LEVEL WAS 21 MG/DL. NO FURTHER INFORMATION WAS AVAILABLE.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
917079 MINIMED QUICK-SET UNO QUICK-SET 60/9 SC1 MECA FPA UNOMEDICAL A/S MMT-397A 6004005 05705244017573

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention| H