VENTED AUTOFEED CHAMBER
Report
- Report Number
- 9611451-2011-00138
- Event Type
- Malfunction
- Date Received
- March 8, 2011
- Date of Event
- December 7, 2010
- Report Date
- February 7, 2011
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BTT
- PMA / PMN Number
- K934140
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). LOT NUMBERS: 100808, 091203, 100402, UNKNOWN. DEVICE MANUFACTURER DATE: 08/08/2010, 12/03/2009, 04/02/2010. METHOD: THE SIX RETURNED COMPLAINT CHAMBERS WERE VISUALLY INSPECTED AND PRESSURE TESTED. ONE COMPLAINT CHAMBER HAD LOT NUMBER 091203, ONE COMPLAINT CHAMBER HAD LOT NUMBER 100402, FOUR COMPLAINT CHAMBERS HAD LOT NUMBER 100808. RESULTS: THE VISUAL INSPECTION REVEALED THAT ALL SIX CHAMBERS WERE DEFORMED NEAR THE CHAMBER BASE. ON ALL SIX COMPLAINT CHAMBERS THERE WERE WHITE MARKS OBSERVED NEAR THE CHAMBER BASE, UP TO THE MAXIMUM WATER FILL LINE. ALL SIX COMPLAINT CHAMBERS WERE PRESSURE TESTED. THE PRESSURE TEST REVEALED THE PRESSURE DROP FOR ALL SIX CHAMBERS TO BE OUTSIDE SPECIFICATION. A LOT CHECK REVEALED: THERE HAS BEEN NO OTHER COMPLAINTS OF THIS NATURE FOR LOT 091203. (B)(4) OTHER SIMILAR COMPLAINT OF THIS NATURE FOR LOT 100402. THERE HAS BEEN NO OTHER COMPLAINTS OF THIS NATURE FOR LOT 100808. CONCLUSION: WE ARE UNABLE TO DETERMINE THE ROOT CAUSE OF THE PROBLEM OBSERVED BY THE CUSTOMER. A FISHER AND PAYKEL HEALTHCARE REPRESENTATIVE IS SCHEDULING A TIME TO VISIT THE HEALTHCARE FACILITY. SHOULD ANY NEW INFORMATION ARISE, WE WILL PROVIDE A FOLLOW-UP REPORT. THE USER INSTRUCTIONS WHICH ACCOMPANY THE MR290 CHAMBER STATE "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT".
(B)(4). THE COMPLAINT DEVICES HAVE BEEN RETURNED TO THE MANUFACTURER FISHER & PAYKEL HEALTHCARE, (B)(4). WE ARE CURRENTLY ENDEAVOURING TO OBTAIN FURTHER INFORMATION WITH REGARD TO THE COMPLAINT AND WILL PROVIDE A FOLLOW-UP REPORT UPON COMPLETION OF OUR INVESTIGATION.
A HOSPITAL IN (B)(6) REPORTED THAT THERE WAS LEAKAGE FROM THE BASE OF THE MR290VX VENTED AUTOFEED CHAMBERS. NO PATIENT CONSEQUENCE WAS REPORTED.
A HOSPITAL IN (B)(6) REPORTED THAT THERE WAS LEAKAGE FROM THE BASE OF THE (B)(4) VENTED AUTOFEED CHAMBERS. NO PATIENT CONSEQENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VENTED AUTOFEED CHAMBER | BTT | BTT | FISHER & PAYKEL HEALTHCARE LTD | MR290VX | SEE. H.10 |