DA VINCI SI SURGICAL SYSTEM
Report
- Report Number
- 2955842-2011-00068
- Event Type
- Other
- Date Received
- March 8, 2011
- Date of Event
- February 7, 2011
- Report Date
- February 7, 2011
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K081137
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC
- Reporter Occupation
- NURSE
Narratives
THE INVESTIGATION CONDUCTED BY THE FIELD SERVICE ENGINEER DETERMINED THAT THE CUSTOMER REPORTED ISSUE WAS ASSOCIATED WITH THE CAMERA CABLE INSTEAD OF THE ENDOSCOPE. THE CAMERA CABLE CONNECTS THE CAMERA TO THE SYSTEM'S VISION CART, WHICH THEN TRANSMITS THE IMAGE TO THE SURGEON'S CONSOLE. THE SYSTEM WAS REPAIRED BY REPLACING THE DEFECTIVE CAMERA CABLE. THE CAMERA CABLE WAS RETURNED AND EVALUATED. ENGINEERING WAS ABLE TO REPLICATE THE REPORTED FAILURE MODE AND CONCLUDED THAT THE ISSUE EXPERIENCED BY THE CUSTOMER WAS A RESULT OF CONDUCTOR DAMAGE. THE DA VINCI SI SURGICAL SYSTEM USER'S MANUAL EXPLICITLY STATES THAT, ENVIRONMENTAL OR EQUIPMENT FAILURES MAY CAUSE THE DA VINCI SI SYSTEM TO BECOME UNAVAILABLE. THE SURGICAL TEAM SHOULD ALWAYS HAVE BACKUP EQUIPMENT AND INSTRUMENTATION AVAILABLE, AND BE PREPARED TO CONVERT TO ALTERNATIVE SURGICAL TECHNIQUES. AS OF MARCH 3, 2011, THERE HAVE BEEN NO REPORTED RECURRENCES OF THE ISSUE AT THIS HOSPITAL.
IT WAS REPORTED THAT PRIOR TO USING THE SYSTEM TO PERFORM ANY SURGICAL TASKS FOR A DA VINCI SI PROSTATECTOMY PROCEDURE, THE SURGICAL STAFF WAS UNABLE TO CALIBRATE THE CAMERA AND THE SYSTEM DIAGNOSTICS GENERATED A MESSAGE INDICATING THAT A POSSIBLE CAUSE WAS A DIRTY SCOPE. THE SITE POWER CYCLED THE SYSTEM, HOWEVER, THIS DID NOT RESOLVE THE ISSUE. THE PATIENT HAD BEEN UNDER ANESTHESIA FOR 20 MINUTES WHEN THE SURGEON MADE THE DECISION TO ABORT THE PLANNED PROCEDURE. NO PATIENT HARM WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DA VINCI SI SURGICAL SYSTEM | ENDOSCOPIC INSTRUMENT CONTROL SYSTEM | NAY | INTUITIVE SURGICAL,INC. | IS3000 6.0P6 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |