DIMENSION (R) CLINICAL CHEMISTRY SYSTEM
Report
- Report Number
- 2517506-2011-00039
- Event Type
- Malfunction
- Date Received
- March 8, 2011
- Date of Event
- February 24, 2011
- Report Date
- February 24, 2011
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
- Product Code
- MMI
- PMA / PMN Number
- K010313
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- OTHER
Narratives
ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED TROPONIN I RESULT IS SAMPLE INTEGRITY. THE IFU FOR DIMENSION CTNI FLEX REAGENT CARTRIDGE CONTAINS THE FOLLOWING INFORMATION: "SPECIMENS SHOULD BE FREE OF PARTICULATE MATTER. TO PREVENT APPEARANCE OF FIBRIN IN SERUM SAMPLES, COMPLETE CLOT FORMATION SHOULD TAKE PLACE BEFORE CENTRIFUGATION." THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
A FALSELY ELEVATED TROPONIN I RESULT WAS OBTAINED ON A PATIENT SAMPLE. THE RESULT WAS NOT REPORTED TO THE PHYSICIAN. THE SAMPLE WAS REPEATED AND A NEGATIVE RESULT WAS OBTAINED AND REPORTED. PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED TROPONIN I RESULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DIMENSION (R) CLINICAL CHEMISTRY SYSTEM | TROPONIN I FLEX(R) REAGENT CARTRIDGE | MMI | SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW | FB1292 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |