FDA Adverse Event Malfunction Summary report: N

DIMENSION (R) CLINICAL CHEMISTRY SYSTEM

MDR report key: 2011007 · Received March 8, 2011

Report

Report Number
2517506-2011-00039
Event Type
Malfunction
Date Received
March 8, 2011
Date of Event
February 24, 2011
Report Date
February 24, 2011
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
Product Code
MMI
PMA / PMN Number
K010313
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED TROPONIN I RESULT IS SAMPLE INTEGRITY. THE IFU FOR DIMENSION CTNI FLEX REAGENT CARTRIDGE CONTAINS THE FOLLOWING INFORMATION: "SPECIMENS SHOULD BE FREE OF PARTICULATE MATTER. TO PREVENT APPEARANCE OF FIBRIN IN SERUM SAMPLES, COMPLETE CLOT FORMATION SHOULD TAKE PLACE BEFORE CENTRIFUGATION." THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.

Description of Event or Problem · 1

A FALSELY ELEVATED TROPONIN I RESULT WAS OBTAINED ON A PATIENT SAMPLE. THE RESULT WAS NOT REPORTED TO THE PHYSICIAN. THE SAMPLE WAS REPEATED AND A NEGATIVE RESULT WAS OBTAINED AND REPORTED. PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED TROPONIN I RESULT.