DIMENSION® CLINICAL CHEMISTRY SYSTEM
Report
- Report Number
- 1226181-2011-00021
- Event Type
- Malfunction
- Date Received
- March 8, 2011
- Date of Event
- February 16, 2011
- Report Date
- February 23, 2011
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC - BROOKFIELD
- Product Code
- JJE
- PMA / PMN Number
- K944093
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED TROPONIN I RESULT WAS HM CHROM CONTAMINATION. A SIEMENS HEALTHCARE DIAGNOSTICS INC. TECHNICAL SERVICE CENTER REPRESENTATIVE DIRECTED THE CUSTOMER TO PERFORM MAINTENANCE. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
THREE FALSELY ELEVATED TROPONIN I RESULTS WERE OBTAINED ON THREE PATIENT SAMPLES IN ONE DAY OF TESTING. THE RESULTS WERE REPORTED TO THE PHYSICIAN WHO QUESTIONED THE RESULTS. THE SAMPLES WERE REPEATED AND NEGATIVE RESULTS WERE OBTAINED AND REPORTED. PATIENT TREATMENTS WERE NOT ALTERED OR PRESCRIBED ON THE BASIS OF THE ELEVATED TROPONIN I RESULTS. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED TROPONIN I RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DIMENSION® CLINICAL CHEMISTRY SYSTEM | CLINICAL CHEMISTRY SYSTEM | JJE | SIEMENS HEALTHCARE DIAGNOSTICS INC - BROOKFIELD | RXL MAX HM |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |