FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CXE VOLUMETRIC INFUSION PUMP SINGLE CHANNEL RECERT

MDR report key: 2010974 · Received March 8, 2011

Report

Report Number
6000001-2011-01682
Event Type
Malfunction
Date Received
March 8, 2011
Date of Event
December 1, 2010
Report Date
December 1, 2010
Manufacturer
BAXTER HEALTHCARE - BUFFALO GROVE
Product Code
FRN
PMA / PMN Number
K063696
Removal / Correction Number
6000001-3/15/05-007-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH MDQ-CAPA-(B)(4). A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH FAILURE CODE 810:04 WAS CONFIRMED IN THE PUMP'S EVENT HISTORY BUT NOT DUPLICATED DURING PRODUCT EVALUATION. THE PUMP'S AIR IN LINE PRINTED CIRCUIT BOARD WAS TESTED AND FOUND TO BE WITHIN SPECIFICATIONS. THEREFORE, NO ASSIGNABLE CAUSE WAS PROVIDED DURING PRODUCT EVALUATION AND NO REPAIR WAS NECESSARY FOR THE REPORTED CONDITION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Description of Event or Problem · 1

THE FACILITY REPORTED A COLLEAGUE INFUSION PUMP WITH FAILURE CODE 810:04. ACCORDING TO THE FACILITY, IT IS UNKNOWN WHEN OR IN WHICH CARE AREA THIS EVENT OCCURRED. UPON REVIEWING THE PUMP'S EVENT HISTORY, BAXTER QUALITY ENGINEERING DISCOVERED THIS EVENT INTERRUPTED DELIVERY. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 6.13.90.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP SINGLE CHANNEL RECERT PUMP, INFUSION FRN BAXTER HEALTHCARE - BUFFALO GROVE

Patients

Seq Age Sex Outcome Treatment
1