FDA Adverse Event Injury Summary report: N

LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM VG

MDR report key: 2010925 · Received March 4, 2011

Report

Report Number
2024601-2011-00118
Event Type
Injury
Date Received
March 4, 2011
Date of Event
November 22, 2010
Report Date
February 3, 2011
Manufacturer
ALLERGAN
Product Code
LTI
PMA / PMN Number
P000008
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TAPER II. (B)(4). VISUAL EXAMINATION OF THE RETURNED DEVICE DETERMINED THE ACCESS PORT TUBING CONNECTOR TO BE A TAPER II. VOMITING AND INTOLERANCE ARE SURGICAL/PHYSIOLOGICAL COMPLICATIONS AND ANALYSIS OF THE DEVICE GENERALLY DOES NOT ASSIST ALLERGAN IN DETERMINING A PROBABLE CAUSE FOR THIS EVENT. ANALYSIS NOTED THE DAMAGED PORT SEPTUM HAD MISSING MATERIAL AND PULLED MATERIAL WITH EVIDENCE OF CORING LIKELY TO HAVE BEEN CAUSED BY THE USE OF A CORING NEEDLE. THE ACCESS PORT HAD NON-PENETRATING NICKS WITH STRIATIONS DESCRIBED AS SURGICAL TOOL SCRATCH-LIKE. THE BAND TUBING AND BUCKLE WERE BROKEN WITH STRIATIONS CONSISTENT WITH A SURGICAL END CUT TO REMOVE THE DEVICE. DEVICE LABELING ADDRESSES THE REPORTED EVENTS OF VOMITING AS FOLLOWS: "NAUSEA AND VOMITING MAY OCCUR, PARTICULARLY IN THE FIRST FEW DAYS AFTER SURGERY AND WHEN THE PATIENT EATS MORE THAN RECOMMENDED."

Description of Event or Problem · 1

HEALTH PROFESSIONAL REPORTED THE EXPLANT OF A LAP-BAND DUE TO "VOMITING AND DIFFICULTY TOLERATING BAND."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM VG LTI ALLERGAN NA 1161710

Patients

Seq Age Sex Outcome Treatment
1 56 YR Required Intervention