LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM VG
Report
- Report Number
- 2024601-2011-00118
- Event Type
- Injury
- Date Received
- March 4, 2011
- Date of Event
- November 22, 2010
- Report Date
- February 3, 2011
- Manufacturer
- ALLERGAN
- Product Code
- LTI
- PMA / PMN Number
- P000008
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
TAPER II. (B)(4). VISUAL EXAMINATION OF THE RETURNED DEVICE DETERMINED THE ACCESS PORT TUBING CONNECTOR TO BE A TAPER II. VOMITING AND INTOLERANCE ARE SURGICAL/PHYSIOLOGICAL COMPLICATIONS AND ANALYSIS OF THE DEVICE GENERALLY DOES NOT ASSIST ALLERGAN IN DETERMINING A PROBABLE CAUSE FOR THIS EVENT. ANALYSIS NOTED THE DAMAGED PORT SEPTUM HAD MISSING MATERIAL AND PULLED MATERIAL WITH EVIDENCE OF CORING LIKELY TO HAVE BEEN CAUSED BY THE USE OF A CORING NEEDLE. THE ACCESS PORT HAD NON-PENETRATING NICKS WITH STRIATIONS DESCRIBED AS SURGICAL TOOL SCRATCH-LIKE. THE BAND TUBING AND BUCKLE WERE BROKEN WITH STRIATIONS CONSISTENT WITH A SURGICAL END CUT TO REMOVE THE DEVICE. DEVICE LABELING ADDRESSES THE REPORTED EVENTS OF VOMITING AS FOLLOWS: "NAUSEA AND VOMITING MAY OCCUR, PARTICULARLY IN THE FIRST FEW DAYS AFTER SURGERY AND WHEN THE PATIENT EATS MORE THAN RECOMMENDED."
HEALTH PROFESSIONAL REPORTED THE EXPLANT OF A LAP-BAND DUE TO "VOMITING AND DIFFICULTY TOLERATING BAND."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM VG | LTI | ALLERGAN | NA | 1161710 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR | Required Intervention |