T15 HEXALOBE, ISO, CANN.
Report
- Report Number
- 1220246-2024-07327
- Event Type
- Malfunction
- Date Received
- August 30, 2024
- Date of Event
- August 9, 2024
- Report Date
- September 30, 2024
- Manufacturer
- ARTHREX, INC.
- Product Code
- LXH
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- 501
Narratives
INVESTIGATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE PROVIDED UPON AVAILABILITY OF ADDITIONAL INFORMATION.
COMPLAINT ALLEGATION IS CONFIRMED. ONE UNPACKAGED AR-8750-03 DRIVER SHAFT BATCH NUMBER: 031833 WAS RECEIVED FOR INVESTIGATION. VISUAL EVALUATION NOTED THAT THE HEX DRIVE GEOMETRY AT THE DISTAL TIP OF THE RETURNED AR-8750-03 DRIVER SHAFT HAD BROKEN OFF. NO PIECES WERE RETURNED FOR INVESTIGATION. FUNCTIONAL TESTING WAS NOT PERFORMED DUE TO THE BROKEN DISTAL TIP OF THE DEVICE. IT WAS NOT INDICATED IF A TORQUE-INDICATING ADAPTER HAD BEEN USED WITH THE DEVICE. THE MOST LIKELY CAUSE FOR THE REPORTED FAILURE CAN BE ATTRIBUTED TO OVER-TORQUING/OVER-ENGAGING THE DRIVER WITHIN THE SCREW HEAD.
ON (B)(6) 2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT (2) AR-8750-03 DRIVER SHAFTS BROKE. THIS OCCURRED DURING A CASE WITH NO EFFECT ON THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED, AND ADDITIONAL INFORMATION HAS BEEN ASKED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1910839 | T15 HEXALOBE, ISO, CANN. | ORTHOPEDIC MANUAL SURG INSTR | LXH | ARTHREX, INC. | T15 HEXALOBE, ISO, CANN. | 031833 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |