HOMECHOICE PRO
Report
- Report Number
- 1423500-2011-02851
- Event Type
- Malfunction
- Date Received
- March 8, 2011
- Date of Event
- February 15, 2011
- Report Date
- February 15, 2011
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K053512
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER; HOWEVER, THE EVALUATION HAS NOT BEEN COMPLETED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE DEVICE EVALUATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB. THE DEVICE'S DEVICE HISTORY RECORD WAS REVIEWED AND NO ISSUES WERE IDENTIFIED THAT MAY HAVE CONTRIBUTED TO THE REPORTED CONDITION. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.
THE HOME PATIENT (HP) CONTACTED BAXTER'S TECHNICAL SERVICE CENTER TO REPORT OVER TEMPERATURE FLUID DELIVERED ON A HOME CHOICE (HC) MACHINE DURING USE. THE HP STATED THE TEMPERATURE OF THE SOLUTION IN THE MANUAL BAG WAS 115 DEGREES FAHRENHEIT. THE BAXTER TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED THAT THE HC DEVICE WAS NOT SUPPOSED TO BE USED HEAT UP MANUAL BAGS. THE TSR EXPLAINED THAT WHEN USING THE HC, THE TEMPERATURE READINGS ARE TAKEN PRIOR TO FILL. THE TSR ASKED THE HP IF THE HC SHOWED TEMPERATURE STABILIZING. THE TSR INITIATED A SWAP OF THE MACHINE. THE HP CONFIRMED TO FINISH THERAPY VIA MANUAL SUPPLIES. THE TSR DISCUSSED THE USE OF CONTINUOUS AMBULATORY PERITONEAL DIALYSIS (CAPD) UNTIL THE NEW MACHINE ARRIVES. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE PRO | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 53 YR |