FDA Adverse Event Malfunction Summary report: N

ACCU-CHEK ® COMPACT TEST DRUM

MDR report key: 2010865 · Received March 8, 2011

Report

Report Number
1823260-2011-01247
Event Type
Malfunction
Date Received
March 8, 2011
Date of Event
February 23, 2011
Report Date
April 7, 2011
Manufacturer
ROCHE DIAGNOSTICS
Product Code
LFR
PMA / PMN Number
K031755
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

IT WAS UNKNOWN IF THE INITIAL REPORTER SENT REPORT TO THE FDA. THIS MEDWATCH REPORT IS FOR THE SUSPECT DEVICE USED IN SYSTEM 1. REFERENCE MEDWATCH REPORT WITH A1 PATIENT IDENTIFIER (B)(6) FOR THE SUSPECT DEVICE USED IN SYSTEM 2.

Description of Event or Problem · 1

REPORTER ALLEGED THE CUSTOMER OBTAINED THE RESULTS OF 159 MG/DL AND 48 MG/DL ON COMPACT PLUS SYSTEM 1 COMPARED BACK TO BACK WITH A RESULT OF 87 MG/DL ON COMPACT PLUS SYSTEM 2 WHEN TESTING WAS PERFORMED WITHIN 10 MINUTES. REPORTER STATED THAT JUST ONE HOUR PRIOR TO OBTAINING THE READINGS, HE HAD TAKEN 3 UNITS OF HUMALOG. NO OTHER ACTIONS WERE REPORTED TAKEN OR TREATMENT RECEIVED. NO ADVERSE EVENT REPORTED. A REQUEST WAS MADE FOR THE RETURN OF THE AFFECTED PRODUCT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ACCU-CHEK ® COMPACT TEST DRUM BLOOD GLUCOSE MONITORING TEST STRIPS LFR ROCHE DIAGNOSTICS NA 207292

Patients

Seq Age Sex Outcome Treatment
1 005 YR HUMALOG| HUMULIN