TALENT ABDOMINAL STENT GRAFT SYSTEM
Report
- Report Number
- 2953200-2011-00587
- Event Type
- Injury
- Date Received
- March 3, 2011
- Date of Event
- February 1, 2011
- Report Date
- February 1, 2011
- Manufacturer
- MEDTRONIC CARDIOVASCULAR GALWAY
- Product Code
- MIH
- PMA / PMN Number
- P070027
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). EVALUATION, RESULTS: ENDOLEAK, SEVERELY ANGULATED NECK. EVALUATION, CONCLUSION: SEVERELY ANGULATED NECK.
A TALENT STENT GRAFT SYSTEM WAS IMPLANTED APPROXIMATELY 2.5 MONTHS AGO, FOR A HIGH RISK PT, FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM. THE AORTA HAD A SHORT NECK WITH A 60-90 DEGREE ANGLE. VESSEL MORPHOLOGY INCLUDES HEAVY NECK AND BILATERAL ILIAC CALCIFICATION AND AN UNK AMOUNT OF THROMBUS. IT WAS REPORTED THAT THE PT WAS ADMITTED VERY SICK TO RECEIVE A LEFT VENTRICULAR ASSIST DEVICE WHEN THE 11 CM ABDOMINAL AORTIC ANEURYSM WAS DISCOVERED. DURING THE TEVAR PROCEDURE, A TYPE I ENDOLEAK APPEARED AND PERSISTED. THE PHYSICIAN PUT IN A TALENT CUFF AND A COMPETITOR'S STENT. A CONTRAST CT AFTER THE PROCEDURE SHOWED THAT THE TYPE I ENDOLEAK WAS RESOLVED. ONE MONTH AGO A PROXIMAL TYPE 1 ENDOLEAK WAS SEEN ON A CONTRAST CT. THE PT WAS NOT A SURGICAL CANDIDATE. TWO WEEKS AGO AN ADDITIONAL TEVAR WAS PERFORMED TO PLACE 2-36 MM CUFFS IN THE VERY ANGULATED NECK. A COMPETITOR'S BALLOON WAS USED. THE PHYSICIAN ASSESSED THAT THE LEAK WAS ANATOMY RELATED NOT DEVICE RELATED. NO CLINICAL SEQUELAE WERE REPORTED, AND THE PT IS FINE. REF, MFR. 2953200-2001-00XXX AND 2953200-2001-00XXX).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TALENT ABDOMINAL STENT GRAFT SYSTEM | MIH | MEDTRONIC CARDIOVASCULAR GALWAY | NA | V00333356 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR | Required Intervention |