INTELLIVUE MX800 PATIENT MONITOR
Report
- Report Number
- 9610816-2024-00562
- Event Type
- Death
- Date Received
- August 30, 2024
- Date of Event
- August 5, 2024
- Report Date
- August 30, 2024
- Manufacturer
- PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
- Product Code
- DSI
- PMA / PMN Number
- K150310
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
- Health Professional
- Yes
Narratives
THE PHILIPS REMOTE CLINICAL SUPPORT (RCS) CONTACTED THE CUSTOMER. IT WAS INDICATED THE X3 WAS DOCKED WITH THE MX800 HOST MONITOR, WHICH WAS NETWORKED WITH THE PIC IX. BOTH MONITORS WERE TESTED BY THE HOSPITAL AND FOUND TO BE WORKING AS EXPECTED. A REVIEW OF THE CLINICAL AUDIT LOGS REVEALED ALARMS WERE GENERATED OVER THE COURSE OF THE 5 HOUR EVENT TIMEFRAME; HOWEVER, THE USER DID NOT ACKNOWLEDGE OR PAUSE ALARMS DURING THAT TIME. THE CUSTOMER INDICATED THERE WAS NEVER ANY ALLEGATION OF DEVICE MALFUNCTION AND THE LOGS WERE REQUESTED DUE TO AN INTERNAL NURSING INVESTIGATION ONLY. THERE WAS NO DELAY IN CARE, THOUGH THE PATIENT WENT TO ICU AND LATER PASSED AWAY. AS THERE WAS NO MALFUNCTION AND NO DELAY IN CARE, THE DEVICE DID NOT CAUSE OR CONTRIBUTE TO THE REPORTED PATIENT DEATH; HOWEVER, THE CAUSE REMAINS UNKNOWN.
IT WAS REPORTED THE CUSTOMER REQUESTED ASSISTANCE IN REVIEWING THE CLINICAL AUDIT LOGS AFTER A PATIENT PASSED AWAY. THE NURSE INDICATED THEY DID NOT HEAR AN ALARM AND THE UNIT MANAGER WAS TRYING TO DETERMINE IF ALARMS WERE ACKNOWLEDGED OR PAUSED AT THE TIME OF THE PATIENT INCIDENT. IT WAS INDICATED THERE WAS NO DELAY AND THE CUSTOMER SIMPLY WANTED TO RESEARCH ALARMS AS THE CUSTOMER WAS INVESTIGATING A STAFF ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1096111 | INTELLIVUE MX800 PATIENT MONITOR | INTELLIVUE MX800 PATIENT MONITOR | DSI | PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH | 865240 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Male | Death |