MAGNESIUM
Report
- Report Number
- 3005094123-2024-00455
- Event Type
- Malfunction
- Date Received
- August 30, 2024
- Date of Event
- August 23, 2024
- Report Date
- September 10, 2024
- Manufacturer
- ABBOTT IRELAND DIAGNOSTICS DIVISION
- Product Code
- JGJ
- UDI-DI
- 00380740176532
- PMA / PMN Number
- K181748
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
AN EVALUATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. SECTION E1 - PHONE NUMBER COMPLETE ENTRY = (B)(6). ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE.
AFTER FURTHER EVALUATION, THE SUSPECT MEDICAL DEVICE WAS CHANGED FROM MAGNESIUM, LIST NUMBER 08P19-34, A.I.D.D LONGFORD, LISNAMUCK, CO. LONGFORD, LONGFORD, IRELAND, N39E932 TO ALINITY C PROCESSING MODULE, LIST NUMBER 03R67-01, ABBOTT LABORATORIES, 1915 HURD DRIVE, IRVING, TX, 75038. MDR NUMBER 3016438761-2024-00534-00 HAS BEEN SUBMITTED AND ALL FURTHER INFORMATION WILL BE DOCUMENTED UNDER THAT MDR NUMBER.
THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY C MAGNESIUM (MG) RESULTS GENERATED FOR A PATIENT SAMPLE. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER¿S REFERENCE RANGE UPPER LIMIT 1.07 MMOL/L): SAMPLE ID (B)(6) INITIAL RESULT = 2.89 MMOL/L, REPEAT RESULT ON ANOTHER INSTRUMENT AC1699 = 0.73 MMOL/L. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY C MAGNESIUM (MG) RESULTS GENERATED FOR A PATIENT SAMPLE. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER¿S REFERENCE RANGE UPPER LIMIT 1.07 MMOL/L): SAMPLE ID (B)(6) INITIAL RESULT = 2.89 MMOL/L, REPEAT RESULT ON ANOTHER INSTRUMENT AC1699 = 0.73 MMOL/L NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1912391 | MAGNESIUM | PHOTOMETRIC METHOD, MAGNESIUM | JGJ | ABBOTT IRELAND DIAGNOSTICS DIVISION | 63239UD00 | 00380740176532 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | ALNTY C PROCESSING MODU, 03R67-01, (B)(6)| ALNTY C PROCESSING MODU, 03R67-01, (B)(6) |