BARD FLAT MESH
Report
- Report Number
- 1213643-2011-00051
- Event Type
- Injury
- Date Received
- February 24, 2011
- Report Date
- February 2, 2011
- Manufacturer
- DAVOL INC., SUB. C.R. BARD, INC.
- Product Code
- FTL
- PMA / PMN Number
- PREAMENDMENT
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
THE INFORMATION PROVIDED INDICATES THAT THE PATIENT DEVELOPED AND WAS TREATED FOR AN INFECTION. THE WARNING SECTION OF THE IFU STATES "IF AN INFECTION DEVELOPS, TREAT THE INFECTION AGGRESSIVELY. THE PROSTHESIS MAY NOT HAVE TO BE REMOVED. AN UNRESOLVED INFECTION, HOWEVER, MAY REQUIRE REMOVAL OF THE PROSTHESIS." IT WAS REPORTED THAT THE PATIENT WAS GIVEN ANTIBIOTICS TO TREAT THE INFECTION. IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT BASED ON THE PROVIDED INFORMATION. WHILE INFECTION IS A KNOWN ADVERSE EVENT THAT IS LISTED IN THE IFU, NO CONCLUSION CAN BE DRAWN AT THIS TIME. A DHR REVIEW HAS BEEN CONDUCTED, INCLUDING STERILITY. ALL PAPERWORK APPEARED COMPLETE AND ACCURATE. NO MANUFACTURING ISSUE WERE IDENTIFIED.
PER PROVIDED MEDICAL RECORDS: ON (B)(6) 2011: PATIENT UNDERWENT LAPAROSCOPIC REPAIR OF LEFT INDIRECT INGUINAL HERNIA WITH INSERTION OF 4 X 6 MESH. ON (B)(6) 2011: PATIENT STARTED ON ORAL ANTIBIOTICS WITH NO IMPROVEMENT WITH REDNESS. FROM (B)(6) 2011: PATIENT PRESENTED WITH PERI-UMBILICAL ABDOMINAL ERYTHEMA, ABDOMINAL PAIN AND DRAINAGE. PATIENT WAS CONSULTED BY INFECTIOUS DISEASE, AND RECEIVED IV ANTIBIOTIC (VANCOMYCIN) AND DEVELOPED A BODY RASH REACTION FROM IT, RED MAN'S SYNDROME. DERMATOLOGY WAS CONSULTED, RECOMMENDED THE RASH FELT MORE LIKE A CONTACT DERMATITIS WITH A MIX OF CELLULITIS. RASH INCREASED OVER SEVERAL DAYS DURING HOSPITAL STAY AND THEN LESSENED. PATIENT DISCHARGED HOME WITH ORAL ANTIBIOTICS AND TOPICAL (STEROID) CREAM TO FOLLOW-UP WITH AN ALLERGIST REGARDING RASH REACTION TO VANCOMYCIN. DISCHARGE DIAGNOSIS: 1) ABDOMINAL WALL CELLULITIS, S/P ABDOMINAL HERNIA REPAIR ON (B)(6) 2011, 2) CONTACT DEMATITIS, 3) RED-MAN SYNDROME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BARD FLAT MESH | FTL | DAVOL INC., SUB. C.R. BARD, INC. | NA | HUUF1489 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 46 YR | Required Intervention |