FDA Adverse Event Malfunction Summary report: N

NC QUANTUM APEX PTCA DILATATION CATHETER

MDR report key: 2010591 · Received March 8, 2011

Report

Report Number
2134265-2011-00550
Event Type
Malfunction
Date Received
March 8, 2011
Date of Event
February 10, 2011
Report Date
February 10, 2011
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LOX
PMA / PMN Number
P860019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AGE AT TIME OF EVENT: (B)(6). DEVICE EVALUATED BY MFR: EXAMINATION OF THE RETURNED DEVICE REVEALED A BALLOON PINHOLE. THE DEVICE WAS PRESSURIZED WITH AN INFLATION DEVICE, WHICH CONFIRMED A PINHOLE ON THE DISTAL END OF THE BALLOON WALL PARALLEL WITH THE DISTAL EDGE OF THE MARKERBAND. INSPECTION OF THE BALLOON PRESENTED NO IRREGULARITIES IN THE BALLOON MATERIAL OR THE RO MARKERS THAT COULD HAVE CONTRIBUTED TO THE DAMAGE. THERE WAS BLOOD AND CONTRAST PRESENT THROUGHOUT THE DISTAL LUMEN AND BALLOON. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTIONAL PROCEDURE A BALLOON RUPTURE OCCURRED. ACCESS WAS GAINED THROUGH THE RADIAL ARTERY. THE 99% STENOSED LESION WAS LOCATED IN THE SEVERELY CALCIFIED AND MODERATELY TORTUOUS LEFT CIRCUMFLEX ARTERY. THE 2.0X8MM NC QUANTUM APEX MONORAIL BALLOON CATHETER WAS ADVANCED TO THE TARGET LESION WHERE THE BALLOON RUPTURED AT 8 ATMS ON THE INITIAL INFLATION. THE PROCEDURE WAS NOT COMPLETED DUE TO ANOTHER REASON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT STATUS IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NC QUANTUM APEX PTCA DILATATION CATHETER CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC - MAPLE GROVE H7493912408200 13751095