FDA Adverse Event Injury Summary report: N

OPTILITE SINGLE-USE HOLMIUM LASER FIBER

MDR report key: 2010561 · Received March 1, 2011

Report

Report Number
1825146-2011-00016
Event Type
Injury
Date Received
March 1, 2011
Date of Event
February 4, 2011
Report Date
March 1, 2011
Manufacturer
COOK UROLOGICAL INC
Product Code
GEX
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

AT THIS TIME, THE DEVICE HAS NOT BEEN RECEIVED FOR EVAL, THE FACILITY HAS BEEN CONTACTED TO INQUIRE WHEN THE PRODUCT WILL BE RETURNED. AS STATED IN THE CUSTOMER STATEMENT, NO HARM TO THE PT OCCURRED HOWEVER, A 2.5CM SEGMENT OF THE FIBER WAS RETRIEVED FROM INSIDE THE PT USING A URETERAL STONE BASKET. DUE TO THE ISSUE WITH THE FIBER, THE PT WILL HAVE TO RETURN AT A LATER DATE TO HAVE THE REMAINDER OF THE STONES FRAGMENTED. AS INFO BECOMES AVAILABLE IT WILL BE FORWARDED.

Description of Event or Problem · 1

THE SURGEON (DR (B)(6)) WAS VAPORIZING A STONE (USING URETERORENOSCOPE P5 FROM OLYMPUS) WITH A NORMAL DEFLEXION (AROUND 45-50 DEGREES), USING ONE OF OUR SINGLE USE LASER FIBERS 273 MICRONS ((B)(4), SERIAL NUMBER (B)(4)). LASER ODYSSEY-30 WAS USED FOR THIS PROCEDURE WITH A 5 HZ AND 1.5 JOULES (=7.5 WATTS). AFTER 30 MINUTES OF VAPORIZATION, SURGEON WAS CLOSE TO FINISH IT, SUDDENLY, WE HAVE SEEN A BIG LIGHT ON THE SCREEN AND A CLEAR DAMAGE APPEARED ON THE RIGHT SIDE OF THE SCREEN. WE'VE PUT THE LASER ON STANDBY AND THE SURGEON REMOVED THE LASER FIBER FROM THE URETEROSCOPE. WE'VE SEEN THAT THE LASER FIBER HAS BROKEN INSIDE THE URETEROSCOPE BECAUSE A PIECE OF APPROXIMATIVELY 2,5 CMS WAS STILL IN THE PT. HE REMOVED ALSO THIS PIECE (LASER FIBER EXTREMITY) WITH AN N-GAGE. THE PT WILL HAVE TO DO AN ADDITIONAL FLEXIBLE URETEROSCOPY TO REFRAGMENT THE STONES WITH A LASER WHICH HAS NOT BEEN FINISHED DURING THE FIRST OPERATION. THERE WERE NO ADVERSE EFFECTS ON THE PT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OPTILITE SINGLE-USE HOLMIUM LASER FIBER GEX LASER INSTRUMENT, SURGICAL GEX COOK UROLOGICAL INC NA U1799963

Patients

Seq Age Sex Outcome Treatment
1 UNK Required Intervention