FDA Adverse Event Malfunction Summary report: N

LENSX LASER SYSTEM

MDR report key: 20105191 · Received August 30, 2024

Report

Report Number
2028159-2024-01283
Event Type
Malfunction
Date Received
August 30, 2024
Date of Event
July 26, 2024
Report Date
November 22, 2024
Manufacturer
ALCON LENSX, INC.
Product Code
OOE
UDI-DI
00380659981623
PMA / PMN Number
K163551
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PRODUCT MANUFACTURING DETAILS UPDATED. ALCON LENSX (SITE #3008772169) IS NO LONGER OPERATIONAL. LENSX MANUFACTURED PRODUCTS ARE MAINTAINED AND INVESTIGATED BY THE ALCON RESEARCH, LTD. IRVINE TECHNOLOGY CENTER (SITE #2028159). THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Additional Manufacturer Narrative · 0

H.3., H.6.: INVESTIGATION, INCLUDING ROOT CAUSE ANALYSIS, IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. ALCON LENSX (SITE #3008772169) IS NO LONGER OPERATIONAL. LENSX MANUFACTURED PRODUCTS ARE MAINTAINED AND INVESTIGATED BY THE ALCON RESEARCH, LTD. IRVINE TECHNOLOGY CENTER SITE #2028159). THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION PROVIDED IN H.3., H.6., AND H.11. A MANUFACTURING DEVICE HISTORY RECORD REVIEW WAS PERFORMED PRIOR TO PRODUCT RELEASE TO ENSURE THAT THE PRODUCT WAS MANUFACTURED IN COMPLIANCE WITH THE DEVICE MASTER RECORD. BASED ON THE ASSESSMENT, THE PRODUCT MET RELEASE CRITERIA. SERVICE ONSITE HISTORY WAS REVIEWED FOR THE SYSTEM. THERE WAS NO SERVICE RECORD (RELEVANT TO THE REPORTED EVENT), FOUND. HOWEVER, THE SYSTEM WAS LAST SERVICED (PRIOR TO THE REPORTED EVENT). THE SYSTEM FOUND TO MEET ALL COSMETIC AND PERFORMANCE STANDARDS, AT THAT TIME. BASED ON THE INFORMATION OBTAINED, THE ROOT CAUSE OF THE REPORTED EVENT IS INCONCLUSIVE. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Description of Event or Problem · 0

A HEALTHCARE PROFESSIONAL REPORTED THAT THE SURGEON OBSERVED THE BUBBLE IN PATIENT INTERFACE OF FLAP AND STROMA WHILE CREATING THE FLAP. THERE WERE ALSO IRREGULARITY UNDERNEATH THE BUBBLE AREA ON STROMA. THE ISSUE WAS REPORTED IN LEFT EYE OF THE PATIENT DURING SURGERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1522813 LENSX LASER SYSTEM OPHTHALMIC FEMTOSECOND LASER OOE ALCON LENSX, INC. 550 NA 00380659981623

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown