LENSX LASER SYSTEM
Report
- Report Number
- 2028159-2024-01283
- Event Type
- Malfunction
- Date Received
- August 30, 2024
- Date of Event
- July 26, 2024
- Report Date
- November 22, 2024
- Manufacturer
- ALCON LENSX, INC.
- Product Code
- OOE
- UDI-DI
- 00380659981623
- PMA / PMN Number
- K163551
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
PRODUCT MANUFACTURING DETAILS UPDATED. ALCON LENSX (SITE #3008772169) IS NO LONGER OPERATIONAL. LENSX MANUFACTURED PRODUCTS ARE MAINTAINED AND INVESTIGATED BY THE ALCON RESEARCH, LTD. IRVINE TECHNOLOGY CENTER (SITE #2028159). THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).
H.3., H.6.: INVESTIGATION, INCLUDING ROOT CAUSE ANALYSIS, IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. ALCON LENSX (SITE #3008772169) IS NO LONGER OPERATIONAL. LENSX MANUFACTURED PRODUCTS ARE MAINTAINED AND INVESTIGATED BY THE ALCON RESEARCH, LTD. IRVINE TECHNOLOGY CENTER SITE #2028159). THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).
ADDITIONAL INFORMATION PROVIDED IN H.3., H.6., AND H.11. A MANUFACTURING DEVICE HISTORY RECORD REVIEW WAS PERFORMED PRIOR TO PRODUCT RELEASE TO ENSURE THAT THE PRODUCT WAS MANUFACTURED IN COMPLIANCE WITH THE DEVICE MASTER RECORD. BASED ON THE ASSESSMENT, THE PRODUCT MET RELEASE CRITERIA. SERVICE ONSITE HISTORY WAS REVIEWED FOR THE SYSTEM. THERE WAS NO SERVICE RECORD (RELEVANT TO THE REPORTED EVENT), FOUND. HOWEVER, THE SYSTEM WAS LAST SERVICED (PRIOR TO THE REPORTED EVENT). THE SYSTEM FOUND TO MEET ALL COSMETIC AND PERFORMANCE STANDARDS, AT THAT TIME. BASED ON THE INFORMATION OBTAINED, THE ROOT CAUSE OF THE REPORTED EVENT IS INCONCLUSIVE. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).
A HEALTHCARE PROFESSIONAL REPORTED THAT THE SURGEON OBSERVED THE BUBBLE IN PATIENT INTERFACE OF FLAP AND STROMA WHILE CREATING THE FLAP. THERE WERE ALSO IRREGULARITY UNDERNEATH THE BUBBLE AREA ON STROMA. THE ISSUE WAS REPORTED IN LEFT EYE OF THE PATIENT DURING SURGERY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1522813 | LENSX LASER SYSTEM | OPHTHALMIC FEMTOSECOND LASER | OOE | ALCON LENSX, INC. | 550 | NA | 00380659981623 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |