FDA Adverse Event
Injury
Summary report: N
VASOVIEW HEMOPRO EVH SYSTEM
MDR report key: 2010485
·
Received March 1, 2011
Report
- Report Number
- 2242352-2011-00102
- Event Type
- Injury
- Date Received
- March 1, 2011
- Date of Event
- January 31, 2011
- Report Date
- January 31, 2011
- Manufacturer
- MAQUET CARDIOVASCULAR, LLC
- Product Code
- GEI
- PMA / PMN Number
- K052274
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WILL REPORTEDLY NOT BE RETURNED TO MAQUET CARDIAC SURGERY FOR INVESTIGATION. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE PRODUCT. THERE WAS NO NONCONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. INTERNAL FILE NUMBER - (B)(4).
Description of Event or Problem · 1
THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, A PIECE OF THE UPPER JAW BOOT TIP BROKE OFF INSIDE THE PATIENT. THE PIECE WAS RECOVERED FROM THE ORIGINAL INCISION. A NEW KIT WAS OPENED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT EFFECTS. THE PRODUCT IS NOT RETURNING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VASOVIEW HEMOPRO EVH SYSTEM | ENDOSCOPIC VESSEL HARVESTING SYSTEM | GEI | MAQUET CARDIOVASCULAR, LLC | VH-3000 | 25025774 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Other |