75MM SELECTABLE NEW TLC
Report
- Report Number
- 3005075853-2011-00934
- Event Type
- Malfunction
- Date Received
- March 8, 2011
- Date of Event
- February 9, 2011
- Report Date
- February 11, 2011
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K092577
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). SWING TAB IN LOCKED POSITION. THE ANALYSIS RESULTS FOUND THAT THE INSTRUMENT WAS RECEIVED IN GOOD VISUAL CONDITION AND WITH A CARTRIDGE RELOAD LOADED IN THE INSTRUMENT. THE CARTRIDGE RELOAD HAD THE SWING TAB IN THE LOCKED POSITION, AND THE PROXIMAL ONE DRIVER UP AND THE REMAINING DRIVERS WERE DOWN WITH STAPLES PRESENT. IT COULD NOT BE DETERMINED IF THE REPORTED INCIDENT WAS THE RESULT OF A PREMATURE LOCKOUT, OR THE RESULT OF AN INTERRUPTED FIRING STROKE. THE SWING TAB WAS RESET AND THEN THE INSTRUMENT WAS TESTED FOR FUNCTIONALITY WITH THE RETURNED CARTRIDGE RELOAD AND IT PERFORMED AS INTENDED. THE DEVICE CLOSED WITHOUT ANY DIFFICULTIES NOTED. WHILE NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT, IT SHOULD BE NOTED THAT 100% INSPECTION IS PERFORMED BY MEANS OF (B)(4) TO INSURE THE SWING TAB IS PRESENT AND UNLOCKED. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A SIGMOIDECTOMY PROCEDURE, ON THE STAPLE LINE CLOSEST TO THE KNIFE AT THE PROXIMAL END ON THE PATIENT SIDE, THE STAPLES WERE NOT FORMED CORRECTLY. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE WITH NO PATIENT CONSEQUENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 75MM SELECTABLE NEW TLC | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | CARTRIDGE |