OMNIPOD INSULIN PUMP
Report
- Report Number
- 3004464228-2011-00063
- Event Type
- Malfunction
- Date Received
- February 11, 2011
- Date of Event
- January 13, 2011
- Report Date
- January 13, 2011
- Manufacturer
- INSULET CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042792
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE POD WILL NOT BE RETURNED TO THE MANUFACTURER FOR EVALUATION. WE ARE UNABLE TO CONFIRM ANY PRODUCT MALFUNCTION. NO CONCLUSION CAN BE REACHED AT THIS TIME. THE CUSTOMER STATED THAT A KINK WAS SEEN IN THE CANNULA. THERE WAS, HOWEVER, NO REPORT OF AN OCCLUSION ALARM. THE POD WOULD HAVE INITIATED AN OCCLUSION ALARM IF THE FLOW OF INSULIN WAS RESTRICTED/PREVENTED BY THE KINK AND RESULTED IN BACK PRESSURE. THE OMNIPOD USER GUIDE INSTRUCTS THE USER TO "CHECK INFUSION SITE FREQUENTLY FOR PROPER CANNULA PLACEMENT". IT ALSO SUGGESTS THAT THE USER SHOULD CHECK THEIR BLOOD GLUCOSE LEVELS FREQUENTLY IN ORDER TO NOTICE AND REACT QUICKLY AND APPROPRIATELY TO ANY ISSUES.
THE CUSTOMER'S MOTHER REPORTED THAT HER SON EXPERIENCED BG LEVELS "UP AS HIGH AS 331MG/DL" AFTER TWO DAYS OF WEARING THE POD. SHE STATED THAT WHEN SHE LOOKED AT THE CANNULA, "IT LOOKED BENT" THOUGH NO ALARM HAD BEEN INITIATED. THE POD WILL NOT BE RETURNED FOR EVALUATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OMNIPOD INSULIN PUMP | PUMP, INFUSION, INSULIN | LZG | INSULET CORPORATION | 11200 | L30390 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |