FDA Adverse Event Malfunction Summary report: N

OMNIPOD INSULIN PUMP

MDR report key: 2010344 · Received February 11, 2011

Report

Report Number
3004464228-2011-00063
Event Type
Malfunction
Date Received
February 11, 2011
Date of Event
January 13, 2011
Report Date
January 13, 2011
Manufacturer
INSULET CORPORATION
Product Code
LZG
PMA / PMN Number
K042792
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE POD WILL NOT BE RETURNED TO THE MANUFACTURER FOR EVALUATION. WE ARE UNABLE TO CONFIRM ANY PRODUCT MALFUNCTION. NO CONCLUSION CAN BE REACHED AT THIS TIME. THE CUSTOMER STATED THAT A KINK WAS SEEN IN THE CANNULA. THERE WAS, HOWEVER, NO REPORT OF AN OCCLUSION ALARM. THE POD WOULD HAVE INITIATED AN OCCLUSION ALARM IF THE FLOW OF INSULIN WAS RESTRICTED/PREVENTED BY THE KINK AND RESULTED IN BACK PRESSURE. THE OMNIPOD USER GUIDE INSTRUCTS THE USER TO "CHECK INFUSION SITE FREQUENTLY FOR PROPER CANNULA PLACEMENT". IT ALSO SUGGESTS THAT THE USER SHOULD CHECK THEIR BLOOD GLUCOSE LEVELS FREQUENTLY IN ORDER TO NOTICE AND REACT QUICKLY AND APPROPRIATELY TO ANY ISSUES.

Description of Event or Problem · 1

THE CUSTOMER'S MOTHER REPORTED THAT HER SON EXPERIENCED BG LEVELS "UP AS HIGH AS 331MG/DL" AFTER TWO DAYS OF WEARING THE POD. SHE STATED THAT WHEN SHE LOOKED AT THE CANNULA, "IT LOOKED BENT" THOUGH NO ALARM HAD BEEN INITIATED. THE POD WILL NOT BE RETURNED FOR EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OMNIPOD INSULIN PUMP PUMP, INFUSION, INSULIN LZG INSULET CORPORATION 11200 L30390

Patients

Seq Age Sex Outcome Treatment
1 Other