FDA Adverse Event
Malfunction
Summary report: N
ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
MDR report key: 2010171
·
Received February 11, 2011
Report
- Report Number
- 2531779-2011-00753
- Event Type
- Malfunction
- Date Received
- February 11, 2011
- Date of Event
- January 11, 2011
- Report Date
- January 11, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Removal / Correction Number
- 2531779-03/24/2010-003-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. ANIMAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. DURING EVALUATION THE FORCE SENSOR PINS WERE FOUND TO BE DAMAGED.
Description of Event or Problem · 1
DURING EVALUATION THE FORCE SENSOR PINS WERE FOUND TO BE DAMAGED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONE TOUCH PING |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR |