ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2011-00791
- Event Type
- Malfunction
- Date Received
- February 11, 2011
- Date of Event
- January 11, 2011
- Report Date
- January 11, 2011
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARS TO BE INTACT WITHOUT ANY OBSERVATION OF PHYSICAL DAMAGE SUCH AS LIFTING OR PEELING. THE ISSUE WITH THE KEYPAD BUTTON MALFUNCTION WAS CONFIRMED THROUGH PRODUCT ANALYSIS. THE "UP/DOWN/OK/CONTRAST" KEYPAD BUTTONS ARE INTERMITTENTLY RESPONDING TO USER INPUTS DURING TESTING. THERE WAS EVIDENCE OF KEYPAD CONTAMINATION/CORROSION FOUND UNDER ALL KEY CONTACTS.
THIS COMPLAINT IS BEING REPORTED DUE TO AN ALLEGED KEYPAD MALFUNCTION. KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THERE WAS NO REPORT OF PRODUCT MISUSE. THE KEYPAD DID NOT APPEAR TO BE PEELED OR TORN. FURTHERMORE, THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ANIMAS 2020 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 15 YR |