FDA Adverse Event Malfunction Summary report: N

REFURBED COLLEAGUE CX VOLUMETRIC INFUSION PUMP SINGLE CHANNE

MDR report key: 2010053 · Received March 8, 2011

Report

Report Number
6000001-2011-01670
Event Type
Malfunction
Date Received
March 8, 2011
Date of Event
February 20, 2011
Report Date
February 23, 2011
Manufacturer
BAXTER HEALTHCARE - BUFFALO GROVE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WHICH OVERINFUSED DURING PATIENT USE WAS NOT CONFIRMED NOR DUPLICATED DURING PRODUCT EVALUATION. DELIVERY ACCURACY TESTS WERE PERFORMED AND FOUND THE PUMP TO BE DELIVERING WITHIN SPECIFICATIONS. THEREFORE, NO ASSIGNABLE CAUSE COULD BE PROVIDED FOR THE REPORTED CONDITION AND NO REPAIR WAS NECESSARY. ADDITIONAL INFORMATION: THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.09.90. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Description of Event or Problem · 1

THE FACILITY REPORTED A COLLEAGUE INFUSION PUMP WHICH OVERINFUSED DURING PATIENT USE IN THE PATIENT RECOVERY DEPARTMENT. ACCORDING TO THE FACILITY REPRESENTATIVE, THE PUMP WAS INFUSING A NORMAL SALINE SOLUTION WHEN THE EVENT OCCURRED. THE PUMP WAS SET TO INFUSE AT A RATE OF 250ML PER HOUR, BUT HAD DELIVERED 500ML WHEN A NURSE RETURNED TO THE ROOM IN APPROXIMATELY AN HOUR. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THE USER INTERFACE MODULE SOFTWARE VERSION OF THIS DEVICE IS CURRENTLY UNKNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 REFURBED COLLEAGUE CX VOLUMETRIC INFUSION PUMP SINGLE CHANNE PUMP, INFUSION FRN BAXTER HEALTHCARE - BUFFALO GROVE

Patients

Seq Age Sex Outcome Treatment
1 SALINE