FDA Adverse Event Injury Summary report: N

PINN MAR+4 NEUT 36IDX52OD

MDR report key: 2010050 · Received February 28, 2011

Report

Report Number
1818910-2011-03130
Event Type
Injury
Date Received
February 28, 2011
Date of Event
January 29, 2011
Report Date
January 29, 2011
Manufacturer
DEPUY ORTHOPAEDICS, INC.
Product Code
LPH
PMA / PMN Number
K001534
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Description of Event or Problem · 1

PT WAS REVISED TO ADDRESS RECURRENT DISLOCATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PINN MAR+4 NEUT 36IDX52OD 87LPH LPH DEPUY ORTHOPAEDICS, INC. NA E5RDL1

Patients

Seq Age Sex Outcome Treatment
1 73 YR Required Intervention