SYNCHRO-14 STRAIGHT 200-35CM
Report
- Report Number
- 3012931345-2024-00156
- Event Type
- Malfunction
- Date Received
- August 29, 2024
- Date of Event
- August 5, 2024
- Report Date
- October 31, 2024
- Manufacturer
- STRYKER NEUROVASCULAR-UTAH-SALT LAKE CITY
- Product Code
- DQX
- UDI-DI
- 07613252186922
- PMA / PMN Number
- K002907
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
D4 GTIN: CORRECTED TO (B)(4). D4 EXPIRATION DATE - ADDED. D9 PRODUCT AVAILABLE TO STRYKER/ RETURNED TO MANUFACTURER ON ¿UPDATED. G4 PMA/510(K) ¿ CORRECTED TO K002907. H3 DEVICE EVALUATED BY MFG ¿ UPDATED. H4 MANUFACTURING DATE ¿ ADDED. THERE ARE CONTROLS IN THE MANUFACTURING PROCESS TO ENSURE THE PRODUCT MET SPECIFICATIONS UPON RELEASE. VISUAL INSPECTION OF THE RETURNED SUBJECT GUIDEWIRE REVEALED THAT PTFE COATING WAS NOTED TO BE PEELED AT 64CM TO 73CM FROM THE PROXIMAL END. THE DISTAL END OF THE SUBJECT GUIDEWIRE WAS NOTED TO BE FRACTURED TO THE NITINOL TUBING WITH THE COIL AND CORE WIRE EXPOSED AT 7CM FROM THE DISTAL END (THIS WAS CONFIRMED TO HAVE OCCURRED POST PROCEDURE). THERE WAS DAMAGE/ ROUGHNESS NOTED TO THE DISTAL HYDROPHILIC COATING IN A NUMBER OF LOCATIONS. FUNCTIONAL TEST WAS PERFORMED AS THE SUBJECT GUIDEWIRE WAS HYDRATED AND WHEN FINGERS WERE RAN ALONG THE LENGTH, ROUGHNESS WAS FELT TO THE HYDROPHILIC COATING. THE REPORTED PTFE COATING PEELING WAS CONFIRMED. THE REPORTED DIFFICULT TO ADVANCE THE SUBJECT GUIDEWIRE COULD NOT BE DUPLICATED DUE TO THE DAMAGE NOTED, HOWEVER, THE DAMAGE NOTED IS CONSISTENT WITH THE REPORTED EVENT. THE REPORTED GUIDEWIRE FRACTURE WAS CONFIRMED. THE REPORTED EVENT IS COVERED IN THE DEVICE DIRECTIONS FOR USE (DFU). AS WELL, THE RISK OF THE REPORTED EVENT IS DOCUMENTED IN THE RISK DOCUMENTATION AND THERE ARE CURRENT CONTROLS TO MITIGATE THE RISK OF THE AS REPORTED EVENT. IT WAS REPORTED THAT FLUSHED AND DELIVERED A SUBJECT GUIDEWIRE AND WHEN MANIPULATE IT THE OPERATOR FOUND THE COATING BETWEEN MIDDLE TO TAIL WAS PEELING AND RESISTANCE WAS ENCOUNTERED WHEN DELIVERED IT TOGETHER WITH SL-10 MICROCATHETER. THE OPERATOR THEN WITHDREW THE SUBJECT GUIDEWIRE AND USED ANOTHER ONE TO CONTINUE THE PROCEDURE SUCCESSFULLY. ADDITIONAL INFORMATION PROVIDED BY THE CUSTOMER INDICATED THAT THE DEVICE WAS PREPARED FOR USE AS PER THE DIRECTIONS FOR USE, THE DEVICE WAS CONFIRMED TO BE IN GOOD CONDITION DURING PREPARATION/PRIOR TO USE ON THE PATIENT, CONTINUOUS FLUSH WAS SET UP AND MAINTAINED THROUGHOUT THE CLINICAL PROCEDURE AND THE PATIENTS ANATOMY WAS MODERATELY TORTUOUS. WHEN PACKING THE SUBJECT GUIDEWIRE ON THE TABLE, THE NURSE UNCAREFULLY BROKEN IT TO FRACTURED. THE SUBJECT GUIDEWIRE DEVICE WAS RETURNED AND IT WAS CONFIRMED THAT THE GUIDEWIRE WAS FRACTURED AND THE PTFE COATING WAS PEELING. THE HYDROPHILIC COATING WAS NOTED TO BE ROUGH AND DAMAGED. THE PEELING NOTED TO THE PTFE COATING IS INDICATIVE OF INTERACTION WITH AN UNDER TIGHTENED TORQUE DEVICE. AN ASSIGNABLE CAUSE OF HANDLING DAMAGE WILL BE ASSIGNED TO THE REPORTED AND ANALYZED ¿ GUIDEWIRE PTFE COATING PEELING¿. IT WAS REPORTED THAT THE GUIDEWIRE WAS BROKEN WHEN BEING PACKED POST PROCEDURE, THEREFORE AN ASSIGNABLE CAUSE OF HANDLING DAMAGE WILL BE ASSIGNED TO THE REPORTED AND ANALYZED ¿ GUIDEWIRE BROKEN/FRACTURED DURING PREPARATION¿. THE DAMAGE NOTED TO THE HYDROPHILIC COATING IS INDICATIVE OF ADVANCING THE HYDRATED SUBJECT GUIDEWIRE AGAINST RESISTANCE, AS WAS REPORTED. IT IS POSSIBLE THAT THE PATIENTS TORTUOUS ANATOMY MAY HAVE CAUSED THE REPORTED DIFFICULTY TO ADVANCE THE SUBJECT GUIDEWIRE AND SUBSEQUENT COATING DAMAGE. AN ASSIGNABLE CAUSE OF PROCEDURAL FACTORS WILL BE ASSIGNED TO THE REPORTED ¿ GUIDEWIRE DIFFICULT TO ADVANCE¿ AND TO THE ANALYZED ¿ GUIDEWIRE HYDROPHILIC COATING SURFACE ROUGH¿, AS THE ISSUE IS ASSOCIATED WITH A PRODUCT THAT MEETS STRYKER DESIGN AND MANUFACTURE SPECIFICATIONS AND WAS USED IN ACCORDING WITH THE DFU BUT DUE TO PROCEDURAL AND/OR ANATOMICAL FACTORS DURING USE, THE PRODUCT PERFORMANCE WAS LIMITED.
IT WAS REPORTED THAT DURING ANEURYSM PROCEDURE, WHEN MANIPULATING THE SUBJECT GUIDEWIRE, THE COATING OF THE SUBJECT GUIDEWIRE WAS FOUND TO BE PEELING BETWEEN THE MIDDLE TO THE TAIL OF THE SUBJECT GUIDEWIRE AND RESISTANCE WAS ENCOUNTERED WHEN DELIVERING THE SUBJECT GUIDEWIRE TOGETHER WITH THE MICROCATHETER. THE SUBJECT GUIDEWIRE WAS REPLACED WITH A NEW GUIDEWIRE AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT CLINICAL CONSEQUENCES TO THE PATIENT.
IT WAS REPORTED THAT DURING ANEURYSM PROCEDURE, WHEN MANIPULATING THE SUBJECT GUIDEWIRE, THE COATING OF THE SUBJECT GUIDEWIRE WAS FOUND TO BE PEELING BETWEEN THE MIDDLE TO THE TAIL OF THE SUBJECT GUIDEWIRE AND RESISTANCE WAS ENCOUNTERED WHEN DELIVERING THE SUBJECT GUIDEWIRE TOGETHER WITH THE MICROCATHETER. THE SUBJECT GUIDEWIRE WAS REPLACED WITH A NEW GUIDEWIRE AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT CLINICAL CONSEQUENCES TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1095953 | SYNCHRO-14 STRAIGHT 200-35CM | WIRE, GUIDE, CATHETER | DQX | STRYKER NEUROVASCULAR-UTAH-SALT LAKE CITY | 0000340273 | 07613252186922 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR | Male | SL-10 MICROCATHETER (STRYKER). |