FDA Adverse Event Malfunction Summary report: N

ADVANTA BED

MDR report key: 2010006 · Received March 2, 2011

Report

Report Number
1824206-2011-01214
Event Type
Malfunction
Date Received
March 2, 2011
Date of Event
February 2, 2011
Report Date
February 2, 2011
Manufacturer
HILL-ROM INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE FOOT HI/LOW WAS NOT FUNCTIONING. HE REPLACED THE FOOT HI/LOW MOTOR TO REPAIR THE BED.

Description of Event or Problem · 1

THE ACCOUNT ALLEGED THE SIDERAIL IS NOT WORKING AND THEY SAW A FLASH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADVANTA BED AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC. 1600

Patients

Seq Age Sex Outcome Treatment
1