FDA Adverse Event Injury Summary report: N

TISSUE-TEK® VIP® 6 AI VACUUM INFILTRATION PROCESSOR; 230 VAC, 50/60 HZ, 6 A

MDR report key: 20098785 · Received August 29, 2024

Report

Report Number
2083544-2024-00011
Event Type
Injury
Date Received
August 29, 2024
Date of Event
July 7, 2024
Report Date
August 27, 2024
Manufacturer
SAKURA SEIKI CO., LTD.
Product Code
IEO
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
EI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

SAKURA FINETEK EUROPE MENTIONED THAT BASED ON THEIR CONVERSATIONS WITH THE CUSTOMER AND LOCAL SAKURA SUPPORT IT HAS BEEN REPORTED THAT THE CUSTOMER'S LAB USES ONE PROGRAM FOR ALL TISSUE SAMPLES. SMALLER TISSUES SUCH AS BIOPSIES AND NEEDLE CORES HAVE NOT BEEN REPORTED TO BE AFFECTED, ONLY LARGER SAMPLES SUCH AS GALL BLADDER AND COLON HAVE BEEN NOTICED TO BE AFFECTED. IT WAS ALSO REPORTED NOT ALL BLOCKS IN ONE CASE WERE AFFECTED. IT IS UNCLEAR WHAT THE EXACT FIXATION TIMES OF THE AFFECTED CASES WERE. BASED ON THE CUSTOMER'S CONVERSATIONS, IN GENERAL SAMPLES ARE TAKEN ON DAY ONE AND FIXED FOR 24 HOURS PRIOR TO DISSECTION AND PROCESSING. MICROTOMY REPORTED NO ISSUES WHEN CUTTING THE CASES, NO CHEMICAL SMELL FROM THE BLOCKS WAS NOTICED. CUSTOMER ALSO MENTIONED THAT THEY DON'T USE A HOT PLATE TO DRY SLIDES. SLIDES ARE PLACED VERTICALLY IN LEICA RACKS AND DRIED IN STAINER OVEN. ON (B)(6) 2024, THE LOCAL FIELD SERVICE ENGINEER VISITED THE CUSTOMER IN IRELAND TO CHECK THE INSTRUMENT. NO MECHANICAL FAULTS WERE DETECTED, AND NO ERRORS WERE LISTED IN THE ERROR LOG. ON THE DATE OF THIS REPORT, THERE IS NOT ENOUGH INFORMATION AVAILABLE YET TO EXCLUDE THE INSTRUMENT FROM (POTENTIALLY) HAVING CONTRIBUTED TO THE SERIOUS INCIDENT. MORE INFORMATION IS BEING GATHERED BY LOCAL SAKURA SUPPORT AND LOGFILES WILL BE TAKEN FROM THE INSTRUMENT FOR ANALYSIS TO CONTINUE THE INVESTIGATION.

Description of Event or Problem · 0

SAKURA FINETEK EUROPE NOTIFIED SAKURA FINETEK USA ON 02- AUG-2024 ABOUT THE INCIDENT THAT OCCURRED BETWEEN (B)(6) 2024 AND (B)(6) 2024, SFE'S LOCAL SAKURA SUPPORT WAS CONTACTED BY THE CUSTOMER IN IRELAND WHO REPORTED THAT A FEW LARGER SAMPLES HAVE SHOWN POOR PROCESSING AND "BURN ARTIFACT", SIMILAR TO CAUTERISATION ARTIFACT WHILE USING TISSUE-TEK® VIP® 6 AI VACUUM INFILTRATION PROCESSOR. SAKURA FINETEK EUROPE ON BECOMING AWARE OF THE INCIDENT ON 26-JUL-2024, WERE INFORMED BY THE CUSTOMER THAT THE ISSUE OF COMPROMISED TISSUE QUALITY MAY HAVE STARTED A FEW WEEKS OR A MONTH AGO BEFORE THE ISSUE WAS REPORTED TO SAKURA FINETEK EUROPE. THE CUSTOMER ALSO MENTIONED THAT THE ISSUE OF POOR PROCESSING IS NOT CONSISTENT WITH EVERY RUN AND NOT ALL BLOCKS ARE IMPACTED. AND THAT THE CUSTOMER ALSO BECAME AWARE OF THE ISSUE BY THE PATHOLOGIST ON THE SAME DAY THEY REPORTED THE INCIDENT TO SAKURA FINETEK EUROPE (B)(6) 2024. PER CUSTOMER, THE TISSUE-TEK® VIP® 6 AI VACUUM INFILTRATION PROCESSOR HAS BEEN FUNCTIONING AS INTENDED AND NO ERRORS HAVE BEEN DISPLAYED ON THE INSTRUMENT DURING PROCESSING. TOTAL TISSUE AND PATIENT IMPACT ARE UNKNOWN AND ON THE DAY OF THE REPORT, ONE CASE HAS BEEN DECLARED AS "UNABLE TO ASSESS TUMOUR DEPTH" IN A COLON RESECTION CASE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1355642 TISSUE-TEK® VIP® 6 AI VACUUM INFILTRATION PROCESSOR; 230 VAC, 50/60 HZ, 6 A VIP6 AI IEO SAKURA SEIKI CO., LTD. 6042

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other