FDA Adverse Event Injury Summary report: N

TRU CC TIB INSERT SIZE 2, 15MM

MDR report key: 20086018 · Received August 28, 2024

Report

Report Number
1038671-2024-03134
Event Type
Injury
Date Received
August 28, 2024
Date of Event
August 12, 2024
Report Date
August 20, 2025
Manufacturer
EXACTECH, INC.
Product Code
JWH
UDI-DI
10885862265890
PMA / PMN Number
K150890
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PENDING INVESTIGATION. D10: 6641890, 02-010-06-0320 - TRU CC FEMORAL SIZE 2 RIGHT. 6931290, 02-012-45-2010 - LGC TIBIAL FIT TRAY CEM SZ 2F/1T. A323354, 02-012-60-1280 - TRU STEM EXT 12MM X 80MM. A207994, 02-012-60-1680 - TRU STEM EXT 16MM X 80MM. 6165557, 208-05-02 - CC DISTAL FEM AUGMENT SZ 2, 5MM. 6314557, 400-40-14 - FLEX OSTEOTOME-FLAT, NAR. A108900, 400-40-14 - FLEX OSTEOTOME-FLAT, NAR.

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL AND/OR CORRECTED INFORMATION. THE REPORTED REVISION DUE TO STIFFNESS CANNOT BE CONCLUSIVELY DETERMINED; HOWEVER, IT MAY BE DUE TO PATIENT FACTORS/ETIOLOGY, COMPONENT POSITIONING, ARTHROFIBROSIS, INFECTION AND/OR IMPLANT SELECTION. THIS CANNOT BE CONFIRMED BECAUSE THE DEVICES WERE NOT AVAILABLE FOR EVALUATION, AND RELEVANT PATIENT INFORMATION, IMAGES, AND RADIOGRAPHS WERE NOT PROVIDED. IF ANY FURTHER INFORMATION IS OBTAINED THAT WOULD CHANGE OR ALTER ANY INFORMATION PROVIDED, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.

Description of Event or Problem · 0

AS REPORTED, APPROXIMATELY ONE YEAR AGO THE PATIENT WAS REVISED FROM COMPETITOR DEVICES TO A CCK. RECENTLY, THE PATIENT CAME IN STIFF, SO WAS DOWNSIZED FROM A 15 CCK INSERT TO A 9 CCK INSERT. THERE WAS NO REPORTED BREAKAGE OF A DEVICE OR SURGICAL DELAY/PROLONGATION. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
179771 TRU CC TIB INSERT SIZE 2, 15MM PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL JWH EXACTECH, INC. 10885862265890

Patients

Seq Age Sex Outcome Treatment
1 NA Female Required Intervention SEE H11.