FDA Adverse Event Injury Summary report: N

STYLE 68 SALINE FILLED BREAST IMPLANT

MDR report key: 20081961 · Received August 27, 2024

Report

Report Number
9617229-2024-19717
Event Type
Injury
Date Received
August 27, 2024
Date of Event
August 6, 2024
Report Date
October 22, 2024
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FWM
PMA / PMN Number
P990074
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

LABORATORY ANALYSIS SUMMARY: THE DEVICE RELATED TO THE REPORTED EVENTS DEFLATION AND VALVE LEAK AND/OR DAMAGE WAS RECEIVED ON SEP 13, 2024 WITH LOT NUMBER 3021413. BASED ON THE PRODUCT ANALYSIS PERFORMED, THE ASSESSMENTS OF THE COMPLAINTS ARE: ¿ DEFLATION: OBSERVED 1 OPENING ASSESSED AS SURGICAL DAMAGE. ¿ VALVE LEAK AND/OR DAMAGE: NO OBSERVED VALVE LEAK. AS PER THE INVESTIGATION PROCEDURE CREASES WAS COMPLETED AND NONE OF THE OBSERVATIONS ARE FOUND TO BE POTENTIALLY RELATED TO THE MANUFACTURING PROCESS, NO FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 0

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: DEFLATION.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED DEFLATION. HEALTHCARE PROFESSIONAL LATER REPORTED HOLE IN VALVE." THIS RECORD IS FOR THE RIGHT SIDE. DEVICE HAS BEEN EXPLANTED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED DEFLATION. HEALTHCARE PROFESSIONAL LATER REPORTED HOLE IN VALVE." THIS RECORD IS FOR THE RIGHT SIDE. DEVICE HAS BEEN EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1479950 STYLE 68 SALINE FILLED BREAST IMPLANT PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM ALLERGAN (COSTA RICA) 3021413

Patients

Seq Age Sex Outcome Treatment
1 39 YR Female Required Intervention