FDA Adverse Event Injury Summary report: N

PROXIMAL HUMERUS, RIGHT, LONG, ÿ 8.5X260MM

MDR report key: 20077747 · Received August 27, 2024

Report

Report Number
0009613350-2024-00407
Event Type
Injury
Date Received
August 27, 2024
Date of Event
August 7, 2024
Report Date
November 27, 2024
Manufacturer
ZIMMER GMBH
Product Code
HSB
UDI-DI
00889024506022
PMA / PMN Number
K200814
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(B)(4). D10. BLUNT TIP SCREW, 4X38MM, ITEM #: 47248603840, LOT #: 3193254. BLUNT TIP SCREW, 4X46MM, ITEM #: 47248604640, LOT #: 3155779. BLUNT TIP SCREW, 4X46MM, ITEM #: 47248604640, LOT #: 3181308. CORTICAL BONE SCREW, 4X24MM, ITEM #: 47248612440, LOT #: 3185173. CORTICAL BONE SCREW, 4X26MM, ITEM #: 47248612640, LOT #: 3193252. PROXIMAL HUMERUS NAIL CAP, 10.5X5MM, ITEM #: 47248801005, LOT #: 3175026. G2. REPORT SOURCE: JAPAN. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS; HOWEVER, AN INVESTIGATION OF THE REPORTED EVENT IS IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETED, A SUPPLEMENTAL MEDWATCH 3500A WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. THE FOLLOWING SECTIONS WERE UPDATED: B4, G3, G6, H2, H3, H6, H10, H11. NO PRODUCT WAS RETURNED OR PICTURES PROVIDED; VISUAL AND DIMENSIONAL EVALUATIONS COULD NOT BE PERFORMED. A REVIEW OF THE DEVICE MANUFACTURING RECORDS CONFIRMED NO ABNORMALITIES OR DEVIATIONS. REVIEW OF COMPLAINT HISTORIES IDENTIFIED ADDITIONAL SIMILAR COMPLAINTS FOR THE REPORTED PRODUCTS AND ADDITIONAL SIMILAR COMPLAINTS FOR THE REPORTED PART AND LOT COMBINATIONS. COMPLAINTS ARE MONITORED THROUGH MONTHLY COMPLAINT REVIEW IN ORDER TO IDENTIFY POTENTIAL ADVERSE TRENDS. IT CAN BE ASSUMED THAT MULTIPLE CONTRIBUTING FACTORS, RELATED TO EITHER PATIENT CONDITION AND BEHAVIOR OR IMPLANTATION PROCEDURE, CONTRIBUTED TO THE MIGRATION OF THE SCREWS. IF AND TO WHAT EXTENT ANY OF THESE ASPECTS MAY HAVE INFLUENCED THE MIGRATION OF THE SCREW REMAINS UNKNOWN. AN ADDITIONAL DEEPER INVESTIGATION WAS PERFORMED WHICH IDENTIFIED THE DESIGN LIMITATION OF THE CORELOCK MECHANISM OF THE AFFIXUS NAIL AS A POTENTIAL CONTRIBUTING FACTOR. AS PART OF THE DEEPER INVESTIGATION, A SCIENTIFIC LITERATURE SEARCH WAS CONDUCTED AND A SYSTEMATIC REVIEW OF THE OUTCOME OF INTRAMEDULLARY NAILING FOR ACUTE PROXIMAL HUMERUS FRACTURE SHOWED THAT SCREW MIGRATION AND PERFORATION INTO THE JOINT IS THE SECOND MOST COMMON COMPLICATION FOLLOWING SECONDARY LOSS OF REDUCTION. IN ADDITION, SCIENTIFIC LITERATURE OF COMPETITOR AND PREDICATE DEVICES WERE ASSESSED. THIS REVIEW HAS SHOWN THAT THE AFFIXUS® NATURAL NAIL® PROXIMAL HUMERAL SYSTEM PERFORMS BETTER AND/OR IN EQUAL RANGE IN COMPARISON WITH LEGALLY MARKETED SIMILAR DEVICES. IN CONCLUSION, AS THE CAUSE MAY BE MULTIFACTORIAL AN EXACT ROOT CAUSE COULD NOT BE IDENTIFIED FOR THE MIGRATION OF THE SCREW. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT AN ANN NAIL SURGERY WAS PERFORMED. AFTER 3 WEEKS FROM THE INITIAL SURGERY, THE SURGEON FOUND THAT ONE OF THE PROXIMAL SCREWS HAD MIGRATED OUT OF THE PROPER POSITION. THE PATIENT COMPLAINED OF PAIN AND REVISION SURGERY WAS PERFORMED TO REMOVE THE MIGRATED SCREW. DILIGENCE IS COMPLETED AN NO FURTHER INFORMATION ON THE REPORTED EVENT HAS BEEN PROVIDED.

Description of Event or Problem · 0

DILIGENCE IS COMPLETED AND NO FURTHER INFORMATION ON THE REPORTED EVENT HAS BEEN PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
179153 PROXIMAL HUMERUS, RIGHT, LONG, ÿ 8.5X260MM PROSTHESIS, TRAUMA HSB ZIMMER GMBH N/A 3187112 00889024506022

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Hospitalization| R SEE H11 NARRATIVE.