FDA Adverse Event Malfunction Summary report: N

C100QT INFANT INCUBATOR

MDR report key: 200750 · Received December 7, 1998

Report

Report Number
2510954-1998-00008
Event Type
Malfunction
Date Received
December 7, 1998
Date of Event
November 9, 1998
Report Date
November 10, 1998
Manufacturer
HILL-ROM AIR-SHIELDS
Product Code
FMZ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

A MONITOR WAS ALARMING AND THE INFANT WAS FOUND ON THE FLOOR AWAKE AND ALERT. IT WAS REPORTED THAT A REAR ACCESS PORT DOOR WAS FOUND OPEN. THE HOSPITAL BIOMEDICAL GROUP REPORTED DAMAGED COMPONENTS WERE FOUND ON THE PORT DOOR LATCH ASSEMBLY, WHICH WERE REPLACED PRIOR TO THE VISIT OF HILL-ROM AIR-SHIELDS SERVICE ENGINEER ON NOVEMBER 10, 1998. NO INJURY RESULTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 C100QT INFANT INCUBATOR INFANT INCUBATOR FMZ HILL-ROM AIR-SHIELDS C100QT NA

Patients

Seq Age Sex Outcome Treatment
1 NA