BD INTIMA-II 24GAX0.75IN PRN SLM NPVC
Report
- Report Number
- 3002601200-2024-00407
- Event Type
- Malfunction
- Date Received
- August 26, 2024
- Date of Event
- July 1, 2024
- Report Date
- September 13, 2024
- Manufacturer
- BD SUZHOU (MDS)
- Product Code
- FOZ
- UDI-DI
- 00382903830787
- PMA / PMN Number
- UNK
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.
1. NO DEFECTIVE SAMPLE AND PHOTOS HAVE BEEN RECEIVED FOR THE COMPLAINT. 2. DHR/BHR REVIEW LOT#4052015 1-THIS BATCH OF PRODUCTS WERE ASSEMBLED AT INTIMA II AUTO LINE 2 IN APRIL 2024, AND PACKAGED AT R240 PACKAGE LINE IN APRIL 2024. WORK ORDER QUANTITY WAS (B)(4) EA. 2-REVIEW THE IN-PROCESS TEST REPORTS AND OUTGOING TEST REPORTS, AND ALL TEST RESULTS MET THE PRODUCT SPECIFICATIONS. 3-REVIEW THE PRODUCTION RECORDS WITH NO NONCONFORMANCE, DEVIATION OR REWORK ACTIVITIES. 4-IN RESPONSE TO THE LEAKAGE AT THE SEPTUM, THE PLANT HAS LAUNCHED AN IN-DEPTH INVESTIGATION TO ALL POTENTIAL CAUSES 3. THE 800MM SIMULATED CLINICAL LEAKAGE TEST WAS CONDUCTED ON THE RETAINED SAMPLES FROM THIS BATCH. IT WAS FOUND THAT A FEW SAMPLES LEAKED AT THE END OF THE SEPTUM AFTER THE NEEDLE TIP WAS INSERTED INTO THE TEST DEVICE, AND THE LEAKAGE STOPPED AFTER THE NEEDLE CORE WAS PULLED OUT. CONCLUSION(S): IN RESPONSE TO THE LEAKAGE AT THE SEPTUM, THE PLANT HAS LAUNCHED AN IN-DEPTH INVESTIGATION TO ALL POTENTIAL CAUSES TO TRACE AND INVESTIGATE ITS ROOT CAUSE.
IT WAS REPORTED THAT BD INTIMA-II 24GAX0.75IN PRN SLM NPVC LEAKED AT SEPTUM AFTER THE INTRAVENOUS INDWELLING NEEDLE WAS PUNCTURED, IT WAS FOUND THAT THE PATIENT'S BLOOD RETURNED AND WAS BLEEDING FROM THE WHITE ISOLATION PLUG.
NO ADDITIONAL INFORMATION PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 277540 | BD INTIMA-II 24GAX0.75IN PRN SLM NPVC | INTRAVASCULAR CATHETER | FOZ | BD SUZHOU (MDS) | 4052015 | 00382903830787 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |