FDA Adverse Event Malfunction Summary report: N

ATLANTIS® ANTERIOR CERVICAL PLATE SYSTEM

MDR report key: 20063812 · Received August 26, 2024

Report

Report Number
1030489-2024-01045
Event Type
Malfunction
Date Received
August 26, 2024
Date of Event
August 8, 2024
Report Date
July 31, 2025
Manufacturer
WARSAW ORTHOPEDICS
Product Code
KWQ
UDI-DI
00613994625137
PMA / PMN Number
K021461
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

NEITHER THE DEVICE NOR FILMS OF APPLICABLE IMAGING STUDIES WERE RETURNED TO THE MANUFACTURER FOR EVALUATION. THEREFORE, WE ARE UNABLE TO DETERMINE THE DEFINITIVE CAUSE OF THE REPORTED EVENT. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

H3: PRODUCT ANALYSIS #828170926:PART # 7200037 : LOT # 0928825W VISUAL INSPECTION CONFIRMED THE MIDDLE LOCKING NUT HAS BROKEN. THE DAMAGE TO THE PLATE IS CONSISTENT WITH OVERLOAD. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING ANTERIOR CERVICAL SPINE SCREW PLATE FIXATION DUE TO ANTERIOR CERVICAL SPINE DOUBLE SEGMENT DEGENERATION.IT WAS REPORTED THAT PLATE BROKE. THE LOCKING RING ON THE ANTERIOR CERVICAL PLATE HAD FALLEN OFF. THIS EVENT OCCURRED BEFORE SURGERY, WHEN THE PATIENT WAS UNDER ANESTHESIA. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1160142 ATLANTIS® ANTERIOR CERVICAL PLATE SYSTEM APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BOD KWQ WARSAW ORTHOPEDICS 7200037 0928825W 00613994625137

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown