FDA Adverse Event Malfunction Summary report: N

POWERED 60 ECHELON +, 340MM SHAFT

MDR report key: 20062999 · Received August 26, 2024

Report

Report Number
3005075853-2024-06272
Event Type
Malfunction
Date Received
August 26, 2024
Date of Event
July 17, 2024
Report Date
August 26, 2024
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
UDI-DI
10705036014607
PMA / PMN Number
K110385
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

(B)(4). DATE SENT: 8/26/2024. D4: BATCH # 864C56. INVESTIGATION SUMMARY THE PRODUCT WAS RETURNED TO ETHICON FOR EVALUATION. VISUAL INSPECTION AND FUNCTIONAL TESTING WERE CONDUCTED ON THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE DETERMINED THAT ONE PSEE60A DEVICE WAS RETURNED WITH NO APPARENT DAMAGE AND WITH NO RELOAD PRESENT. THE DEVICE WAS RETURNED WITH A NON-WORKING FIRING MECHANISM. NO FURTHER FUNCTIONAL TEST COULD BE PERFORMED DUE TO THE CONDITION OF THE DEVICE. IN ORDER TO VERIFY THE CONDITION OF THE INTERNAL COMPONENTS, THE DEVICE WAS DISASSEMBLED. UPON DISASSEMBLING, THE PCB BOARD WAS NOTED TO BE NON-FUNCTIONAL. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE PCB BOARD TO BE DAMAGED. DEVICE HISTORY REVIEW A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE BATCH NUMBER 864C56, AND NO NON-CONFORMANCES WERE IDENTIFIED.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING THE LAPAROSCOPIC LOBECTOMY SURGERY, COULD NOT FIRE WITHOUT MOTOR SOUND ON THE 1TH FIRING. THE SURGEON REMOVED THE BATTERY AND ROTATED FOR 180 DEGREE AND REINSERTED THE BATTERY. THE DEVICE STILL COULD NOT FIRE AND WITHOUT MOTOR SOUND. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
276335 POWERED 60 ECHELON +, 340MM SHAFT STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. C9DE9L 10705036014607

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown