AUTOSOFT XC
Report
- Report Number
- 3003442380-2024-22309
- Event Type
- Malfunction
- Date Received
- August 23, 2024
- Date of Event
- May 19, 2024
- Report Date
- July 26, 2024
- Manufacturer
- UNOMEDICAL A/S
- Product Code
- FPA
- UDI-DI
- 05705244016712
- PMA / PMN Number
- K032854
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
INITIAL AND FINAL MDR 1950568- MDR 3003442380-2024-22309- DEVICE 1 OF 2. SINCE NO LOT NUMBER IS AVAILABLE, A DETAILED INVESTIGATION, TESTS ON REFERENCE SAMPLES OR BATCH REVIEW CANNOT BE CONDUCTED. THEREFORE, THIS COMPLAINT WILL BE CLOSED, THIS ISSUE WILL BE MONITORED THROUGH PMS PRODUCT TRENDS AND MALFUNCTION. IF ANY TRENDS PICKED UP, THIS WILL FLAG ON THE TRIPS AND ALERTS ACCORDING TO THE OMQR PROCEDURE.
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT EXPERIENCED CANNULA ISSUES WITH TWO INFUSION SETS. THE CANNULAS WERE KINKED. THE EVENT OCCURRED ON 19-MAY-2024. PATIENT NOTICED SYMPTOMS WITHIN 3 OR MORE HOURS AFTER INSERTION. INFUSION SET WAS USED FOR LESS THAN 24 HOURS. THE SITE OF INSERTION WAS UPPER BUTTOCKS, ABDOMEN, HIPS, ARM. PATIENT REGULARLY ROTATED THE SITE LOCATION. PATIENT CONFIRMED THAT THE INTRODUCER NEEDLE WAS AHEAD OF THE CANNULA PRIOR TO INSERTION. BLOOD GLUCOSE LEVEL WAS 600 MG/DL AT THE TIME OF THE EVENT. THEREFORE, PATIENT TOOK CORRECTION INJECTION VIA MDI FOR THE TREATMENT. PATIENT REPLACED THE INFUSION SET AND SUCCESSFULLY RESUMED INSULIN DELIVERIES. COMPANY DO NOT SEE BENT/KINKING AS BEING RELATED TO HUMAN FACTORS, BUT RATHER AS A TRAINING ISSUE INCLUDING CORRECT CHOICES OF INSERTION SITES AND INFUSION SETS AND CANNULA LENGTH. FURTHERMORE, THE SOFT CANNULA IS A FLEXIBLE MATERIAL THAT DURING USE AND UPON REMOVAL CAN BEND SLIGHTLY. NO FURTHER INFORMATION WAS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1159636 | AUTOSOFT XC | UNO INSET I 60/6 BLUE TCAP 10PK INT | FPA | UNOMEDICAL A/S | 1001683 | UNKNOWN | 05705244016712 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 15 YR | Female |