FDA Adverse Event Injury Summary report: N

SIGN IM NAIL

MDR report key: 20048203 · Received August 23, 2024

Report

Report Number
3034525-2024-00149
Event Type
Injury
Date Received
August 23, 2024
Date of Event
July 29, 2024
Report Date
August 8, 2024
Manufacturer
SIGN FRACTURE CARE INTERNATIONAL
Product Code
HSB
PMA / PMN Number
K022632
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
BG
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A PRODUCT INVESTIGATION WAS PERFORMED FOR THIS DEVICE. THE ACTUAL DEVICE WAS NOT RETURNED TO THE MANUFACTURER FOR EVALUATION. THE ROOT CAUSE OF THE NON-UNION IS UNDETERMINED. THERE IS NO WAY TO PREDICT A NON-UNION OR FAILURE TO HEAL. THE RADIOGRAPHIC AND CLINICAL DATA WERE REVIEWED BY A SIGN ORTHOPEDIC SURGEON. THIS FAILURE DOES NOT INDICATE A DEFECT IN THE PRODUCT. A MINIMAL RISK IS ASSOCIATED WITH THIS FAILURE. SIGN FRACTURE CARE INTERNATIONAL CONTINUES TO MONITOR THESE EVENTS AS PART OF OUR POST MARKET SURVEILANCE ACTIVITIES.

Description of Event or Problem · 0

WE BECAME AWARE ON (B)(6)2024 THAT A SIGN IM NAIL IMPLANTED TO REPAIR A FRACTURE WAS REPLACED DUE TO A NON-UNION. THE IM NAIL WAS REPLACED WITH A 12MM X 340MM STANDARD IM NAIL PER THE SIGN TECHNIQUE MANUAL. SURGEON COMMENT: "-GENEROUS ILIAC BONE GRAFT TO FRACTURE - SEVERELY OSTEOPOROTIC DISTAL TIB WOULD NOT HOLD SCREWS - #1 PDS CERCLAGE X 4 AROUND BUTTERFLY FRAGMENT + IBG".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1581919 SIGN IM NAIL INTRAMEDULLARY FIXATION ROD HSB SIGN FRACTURE CARE INTERNATIONAL

Patients

Seq Age Sex Outcome Treatment
1 50 YR Female Required Intervention| H