FDA Adverse Event Malfunction Summary report: N

FLUSHING PUMP OFP-2 (JP)

MDR report key: 20044311 · Received August 22, 2024

Report

Report Number
9611174-2024-01447
Event Type
Malfunction
Date Received
August 22, 2024
Date of Event
August 9, 2024
Report Date
September 18, 2024
Manufacturer
KEYMED (MEDICAL AND INDUSTRIAL EQUIPMENT) LTD.
Product Code
FEQ
PMA / PMN Number
K100899
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

E2/E3: NO INFORMATION AVAILABLE FOR THE OCCUPATION OF THE INITIAL REPORTER OR WHETHER THEY ARE A HEALTHCARE PROFESSIONAL. THE INVESTIGATION IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETED OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE RESULTS OF THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION AND THE REPORTED FAILURE " FLOW RATE ADJUSTMENT NOT POSSIBLE" WAS CONFIRMED. IN ADDITION, THE FOLLOWING REPORTABLE MALFUNCTIONS WERE IDENTIFIED DURING THE DEVICE EVALUATION: FLOW RATE CONTROLS BUTTON DID NOT RESPOND, THE POWER INDICATOR DID NOT ILLUMINATE. A DEFINITIVE ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THE FOLLOWING LED TO THE MALFUNCTION: COMPONENT FAILURE SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE SUBJECT DEVICE'S FLOW RATE ADJUSTMENT WAS NOT POSSIBLE THERE WERE NO REPORTS OF PATIENT HARM.

Description of Event or Problem · 0

ADDITIONAL INFORMATION WAS PROVIDED BY THE CUSTOMER: THE ISSUE WAS IDENTIFIED DURING THE BEGINNING OF THERAPEUTIC ENDOSCOPIC SUBMUCOSAL DISSECTION PROCEDURE, THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1528173 FLUSHING PUMP OFP-2 (JP) OLYMPUS FLUSHING PUMP FEQ KEYMED (MEDICAL AND INDUSTRIAL EQUIPMENT) LTD. K10001142

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown