FDA Adverse Event Malfunction Summary report: N

FG 6301 DUAL CHNL VOL INF PUMP2M8064 SERVICE

MDR report key: 2003959 · Received March 1, 2011

Report

Report Number
6000001-2011-01457
Event Type
Malfunction
Date Received
March 1, 2011
Date of Event
September 12, 2010
Report Date
February 24, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915523
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE CONDITION FOR A FLOGARD INFUSION PUMP WITH A FALSE OCCLUSION ALARM WAS FOUND IN THE PUMP'S EVENT HISTORY BUT NOT DUPLICATED DURING PRODUCT EVALUATION. THEREFORE, NO ASSIGNABLE CAUSE COULD BE PROVIDED FOR THIS CONDITION AND NO REPAIR WAS NECESSARY. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

DURING A REVIEW OF THE PUMP'S EVENT HISTORY BY BAXTER PERSONNEL, A FLO-GARD INFUSION PUMP WAS FOUND TO HAVE EXPERIENCED F3 TWO TIMES ON THE REPORTED OCCURENCE DATE. DURING DEVICE EVALUATION, THE UPPER HINGE STOP AND LOWER HINGE STOP WERE FOUND BROKEN ON THE DOOR OF THIS DEVICE, INDICATING THAT THESE OCCURRENCES OF F3 REPRESENT FALSE OCCLUSION ALARMS. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FG 6301 DUAL CHNL VOL INF PUMP2M8064 SERVICE PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1