FDA Adverse Event Injury Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 20039560 · Received August 22, 2024

Report

Report Number
3008642652-2024-08499
Event Type
Injury
Date Received
August 22, 2024
Date of Event
August 5, 2024
Report Date
August 19, 2024
Manufacturer
ZOLL MANUFACTURING CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

BASED ON THE DATA AVAILABLE AT THIS TIME, THERE IS NO INDICATION OF A DEVICE MALFUNCTION CAUSING OR CONTRIBUTING TO THE INAPPROPRIATE TREATMENT. INAPPROPRIATE DEFIBRILLATIONS ARE AN ANTICIPATED RISK ASSOCIATED WITH THE USE OF THE LIFEVEST. PATIENTS ARE INSTRUCTED THROUGH ALARMS, VOICE MESSAGES, IFU, AND TRAINING TO PRESS THE RESPONSE BUTTONS TO PREVENT AN INAPPROPRIATE DEFIBRILLATION. THE CURRENT COMMERCIAL INAPPROPRIATE DEFIBRILLATION RATE IS CONSISTENT WITH THE OBSERVED RATE DURING THE PIVOTAL CLINICAL TRIAL G960083 (0.69% PER PATIENT-MONTH WITH 90% CONFIDENCE). A SUMMARY OF THE SAFETY AND EFFECTIVENESS DATA (SSED), INCLUDING THE INAPPROPRIATE DEFIBRILLATION SAFETY OBJECTIVE SUPPORTING FDA'S APPROVAL OF THE LIFEVEST, CAN BE FOUND AT HTTP://WWW.ACCESSDATA.FDA.GOV/CDRH_DOCS/PDF/P010030B.PDF. THE LIFEVEST DETECTION ALGORITHM COMPLIES WITH IEC 60601-2-4 PERFORMANCE REQUIREMENTS FOR SENSITIVITY AND SPECIFICITY.

Description of Event or Problem · 0

A PATIENT RECEIVED AN INAPPROPRIATE SHOCK. DOUBLE COUNTING CONTRIBUTED TO THE FALSE DETECTION. THE RESPONSE BUTTONS WERE NOT PRESSED DURING THE EVENT. THE PATIENT WENT TO THE HOSPITAL FOR FURTHER EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
27016 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown