FDA Adverse Event
Malfunction
Summary report: N
NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE
MDR report key: 2003867
·
Received February 7, 2011
Report
- Report Number
- 1811755-2011-00363
- Event Type
- Malfunction
- Date Received
- February 7, 2011
- Date of Event
- August 4, 2010
- Report Date
- August 4, 2010
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- FYD
- PMA / PMN Number
- K012991
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE FIELD SVC TECH INSPECTED THE UNIT AND FOUND A HUBBLE LOCK STYLE END ON THE PLUG. HE INSPECTED THE INSIDE OF THE PLUG AND FOUND THAT THE ENDS OF IT WERE BURNT. THE POWER PLUG WAS REPLACED, THE UNIT PASSED ALL REQUIRED TESTS, AND IT WAS PUT BACK INTO SVC.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE POWER CORD DID NOT HAVE POWER. THE FIELD SVC TECH FOUND THE INSIDE OF THE PLUG WAS BURNT. THERE WERE NO ADVERSE CONSEQUENCES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE | AIR-HANDLING APPARATUS FOR A SURGICAL OPERATING RO | FYD | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |