FDA Adverse Event Malfunction Summary report: N

NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE

MDR report key: 2003867 · Received February 7, 2011

Report

Report Number
1811755-2011-00363
Event Type
Malfunction
Date Received
February 7, 2011
Date of Event
August 4, 2010
Report Date
August 4, 2010
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
FYD
PMA / PMN Number
K012991
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SVC TECH INSPECTED THE UNIT AND FOUND A HUBBLE LOCK STYLE END ON THE PLUG. HE INSPECTED THE INSIDE OF THE PLUG AND FOUND THAT THE ENDS OF IT WERE BURNT. THE POWER PLUG WAS REPLACED, THE UNIT PASSED ALL REQUIRED TESTS, AND IT WAS PUT BACK INTO SVC.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE POWER CORD DID NOT HAVE POWER. THE FIELD SVC TECH FOUND THE INSIDE OF THE PLUG WAS BURNT. THERE WERE NO ADVERSE CONSEQUENCES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE AIR-HANDLING APPARATUS FOR A SURGICAL OPERATING RO FYD STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK