LIFEPAK 20 DEFIBRILLATOR/MONITOR
MDR report key: 2003764
·
Received February 7, 2011
Report
- Report Number
- 3015876-2011-00115
- Event Type
- Malfunction
- Date Received
- February 7, 2011
- Date of Event
- January 11, 2011
- Report Date
- January 11, 2011
- Manufacturer
- PHYSIO-CONTROL, INC
- Product Code
- LDD
- PMA / PMN Number
- K012274
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
(B)(4): PHYSIO-CONTROL REPLACED THE PROGRAMMED SYSTEM CONTROLLER PCB AND USER INTERFACE PCB ASSEMBLIES TO OBSERVE PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS THEN RETURNED TO THE CUSTOMER FOR USE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE DEVICE WAS NOT FUNCTIONING. PHYSIO-CONTROL FOUND THAT ALL THE BUTTONS EXCEPT THE POWER ON/OFF BUTTON WERE INOPERATIVE, A LOCK UP FAILURE. THERE WAS NO PATIENT USE ASSOCIATED WITH THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEPAK 20 DEFIBRILLATOR/MONITOR | LDD | PHYSIO-CONTROL, INC | 20 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |