FDA Adverse Event Malfunction Summary report: N

LIFEPAK 20 DEFIBRILLATOR/MONITOR

MDR report key: 2003764 · Received February 7, 2011

Report

Report Number
3015876-2011-00115
Event Type
Malfunction
Date Received
February 7, 2011
Date of Event
January 11, 2011
Report Date
January 11, 2011
Manufacturer
PHYSIO-CONTROL, INC
Product Code
LDD
PMA / PMN Number
K012274
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

(B)(4): PHYSIO-CONTROL REPLACED THE PROGRAMMED SYSTEM CONTROLLER PCB AND USER INTERFACE PCB ASSEMBLIES TO OBSERVE PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS THEN RETURNED TO THE CUSTOMER FOR USE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE DEVICE WAS NOT FUNCTIONING. PHYSIO-CONTROL FOUND THAT ALL THE BUTTONS EXCEPT THE POWER ON/OFF BUTTON WERE INOPERATIVE, A LOCK UP FAILURE. THERE WAS NO PATIENT USE ASSOCIATED WITH THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEPAK 20 DEFIBRILLATOR/MONITOR LDD PHYSIO-CONTROL, INC 20 NA

Patients

Seq Age Sex Outcome Treatment
1 NA