FDA Adverse Event Injury Summary report: N

PROCARE

MDR report key: 20037459 · Received August 22, 2024

Report

Report Number
9616086-2024-00057
Event Type
Injury
Date Received
August 22, 2024
Date of Event
July 19, 2023
Report Date
August 21, 2024
Manufacturer
DJO LLC
Product Code
KYS
PMA / PMN Number
E317265
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

IT WAS REPORTED THAT ONE WEEK AFTER WEARING THE POST OP SHOE A SPOT UNDER SECOND TOE RIGHT FOOT. I RETURNED TO DR. EXAMINED THE SHOE AND SAW A PIECE OF METAL UNDER THE CUSHION WAS COMING THROUGH AND RUBBING MY FOOT. THE PATIENT IS DIABETIC, SO THE DR STARTED TREATING THE WOUND, AFTER 3 WEEKS HAD GOTTEN LARGER. THE PATIENT RECEIVED TREATMENT FOR ANOTHER 2 MONTHS TRYING TO GET THE WOUND TO HEAL. ON (B)(6) 2023, THE PATIENT WENT SEPTIC FROM THE WOUND ORAL ANTIBIOTICS WHERE NOT WORKING. THE PATIENT WAS ADMITTED TO THE HOSPITAL ON (B)(6), 2023, WHERE AFTER IV ANTIBIOTICS THEY DECIDED TO AMPUTATE THE PATIENTS TOE. THE PATIENT WENT HOME AND WAS READMITTED INTO THE HOSPITAL ON (B)(6)2023, TO REMOVE MORE BONE DUE TO NOT HEALING AT THAT TIME. THE PATIENT WAS SENT HOME THIS TIME WITH A PICC LINE FOR ANTIBIOTICS AND RECEIVED IV ANTIBIOTICS UNTIL (B)(6) 2024. THE DEVICE WAS NOT RETURNED TO ENOVIS FOR EVALUATION, THE ROOT CAUSE COULD NOT BE ESTABLISHED. IF ADDITIONAL INFORMATION REGARDING THE REPORTED EVENT IS SUBMITTED AT A FUTURE DATE, THIS INVESTIGATION WILL BE RE-EVALUATED.

Description of Event or Problem · 0

IT WAS REPORTED THAT ONE WEEK AFTER WEARING THE POST OP SHOE A SPOT UNDER SECOND TOE RIGHT FOOT. I RETURNED TO DR. EXAMINED THE SHOE AND SAW A PIECE OF METAL UNDER THE CUSHION WAS COMING THROUGH AND RUBBING MY FOOT. THE PATIENT IS DIABETIC, SO THE DR STARTED TREATING THE WOUND, AFTER 3 WEEKS HAD GOTTEN LARGER. THE PATIENT RECEIVED TREATMENT FOR ANOTHER 2 MONTHS TRYING TO GET THE WOUND TO HEAL. ON (B)(6) 2023, THE PATIENT WENT SEPTIC FROM THE WOUND ORAL ANTIBIOTICS WHERE NOT WORKING. THE PATIENT WAS ADMITTED TO THE HOSPITAL ON (B)(6) 2023, WHERE AFTER IV ANTIBIOTICS THEY DECIDED TO AMPUTATE THE PATIENTS TOE. THE PATIENT WENT HOME AND WAS READMITTED INTO THE HOSPITAL ON (B)(6) 2023, TO REMOVE MORE BONE DUE TO NOT HEALING AT THAT TIME. THE PATIENT WAS SENT HOME THIS TIME WITH A PICC LINE FOR ANTIBIOTICS AND RECEIVED IV ANTIBIOTICS UNTIL (B)(6) 2024.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
968695 PROCARE SQUARED TOE POST OP SHOE, L KYS DJO LLC UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 NA Male Disability