FDA Adverse Event Injury Summary report: N

VASOVIEW HEMOPRO EVH SYSTEM

MDR report key: 2003710 · Received February 23, 2011

Report

Report Number
2242352-2011-00085
Event Type
Injury
Date Received
February 23, 2011
Date of Event
January 25, 2011
Report Date
January 25, 2011
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
GEI
PMA / PMN Number
K052274
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION SUMMARY: A VISUAL INSPECTION REVEALED THAT THE TIP OF THE COLD JAW BOOT HAD DETACHED AND WAS RECEIVED SEPARATELY. THE JAWS WERE SOMEWHAT BURNT. THERE WAS SOME EVIDENCE OF BLOOD. BASED UPON THE VISUAL OBSERVATIONS, THE REPORTED COMPLAINT FOR "JAW BOOT BROKE OFF" WAS CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE PRODUCT. THERE WAS NO NONCONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. INTERNAL FILE NUMBER - (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE SILICON JAW BOOT BROKE OFF. THE PIECE WAS RETRIEVED WITH NO OTHER PATIENT EFFECTS REPORTED. A NEW KIT WAS USED TO COMPLETE THE PROCEDURE. THE PRODUCT WAS RETURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VASOVIEW HEMOPRO EVH SYSTEM ENDOSCOPIC VESSEL HARVESTING SYSTEM GEI MAQUET CARDIOVASCULAR, LLC VH-3000 25024926

Patients

Seq Age Sex Outcome Treatment
1 NA Other