FDA Adverse Event Injury Summary report: N

STRATA II, ADJUSTABLE DELTA VALVE, SMALL

MDR report key: 2003630 · Received February 23, 2011

Report

Report Number
2021898-2011-00043
Event Type
Injury
Date Received
February 23, 2011
Date of Event
January 24, 2011
Report Date
January 24, 2011
Manufacturer
MEDTRONIC NEUROSURGERY
Product Code
JXG
PMA / PMN Number
K042465
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE RETURNED VALVE WAS PATENT, PASSED LEAK TESTING AND MET ALL SPECIFICATIONS FOR PRESSURE-FLOW AND PREIMPLANTATION TESTING AT 0 CM HYDROSTATIC PRESSURE. HOWEVER, IT DID NOT MEET THE REQUIREMENTS FOR REFLUX OR SIPHON TESTING, WHICH PRECLUDED MEASURING OF PRESSURE-FLOW AT -50 CM HYDROSTATIC PRESSURE. CRYSTALLINE DEBRIS WAS FOUND ON THE EXTERIOR AND INTERIOR OF THE VALVE, PARTICULARLY ON THE DIAPHRAGMS OF THE DELTA CHAMBER. DEBRIS WITHIN THE VALVE MAY INTERFERE WITH THE ANTI-SIPHON DEVICE, RESULTING IN SIPHONING. ALL PERFORMANCE LEVELS COULD BE SET WITH A SINGLE ATTEMPT. A DISSECTION OF THE ADJUSTMENT MECHANISM SHOWED NO ANOMALIES. IT IS UNK WHAT MAY HAVE CAUSED THE REPORTED LEVEL CHANGE. STRATA PRODUCTS ARE DESIGNED TO BE ADJUSTED BY A STRONG MAGNET. INADVERTENT ADJUSTMENTS BY EXTERNAL MAGNETS ARE A KNOWN COMPLICATION WITH ADJUSTABLE VALVES. THE INSTRUCTIONS FOR USE THAT ACCOMPANIES THE PRODUCT CAUTIONS THE DEVICES KNOWN TO CONTAIN MAGNETS SHOULD BE KEPT AWAY FROM THE IMMEDIATE VALVE IMPLANT LOCATION, AS THEY MAY HAVE AN EFFECT ON THE PERFORMANCE LEVEL SETTING OF THE STRATA-TYPE VALVE. ALL MAGNETS HAVE AN EXPONENTIALLY DECREASING EFFECT ON THE VALVE THE FURTHER AWAY THEY ARE LOCATED. WHILE IT IS IMPOSSIBLE TO QUANTIFY ALL THE VARIOUS MAGNETIC FIELDS IN OUR ENVIRONMENT, THE VAST MAJORITY WILL NOT IMPACT THE VALVE AS LONG AS DISTANCE FROM THE VALVE IMPLANT IS MAINTAINED. A REVIEW OF THE MFG RECORDS WAS NOT POSSIBLE AS NO LOT NUMBER WAS PROVIDED. ALL OF OUR VALVES ARE 100% TESTED AT THE TIME OF MFR.

Description of Event or Problem · 1

IT WAS REPORTED TO MEDTRONIC NEUROSURGERY THAT THE VALVE WAS CHANGING SETTINGS WITH NO OUTSIDE INTERFERENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STRATA II, ADJUSTABLE DELTA VALVE, SMALL JXG MEDTRONIC NEUROSURGERY NA UNK

Patients

Seq Age Sex Outcome Treatment
1 UNK Hospitalization| R